High temporal-resolution scanning transmission electron microscopy using sparse-serpentine scan pathways

Scanning transmission electron microscopy (STEM) provides structural analysis with sub-angstrom resolution. But the pixel-by-pixel scanning process is a limiting factor in acquiring high-speed data. Different strategies have been implemented to increase scanning speeds while at the same time minimiz...

Full description

Saved in:
Bibliographic Details
Published inScientific reports Vol. 11; no. 1; p. 22722
Main Authors Ortega, Eduardo, Nicholls, Daniel, Browning, Nigel D., de Jonge, Niels
Format Journal Article
LanguageEnglish
Published London Nature Publishing Group UK 22.11.2021
Nature Publishing Group
Nature Portfolio
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Scanning transmission electron microscopy (STEM) provides structural analysis with sub-angstrom resolution. But the pixel-by-pixel scanning process is a limiting factor in acquiring high-speed data. Different strategies have been implemented to increase scanning speeds while at the same time minimizing beam damage via optimizing the scanning strategy. Here, we achieve the highest possible scanning speed by eliminating the image acquisition dead time induced by the beam flyback time combined with reducing the amount of scanning pixels via sparse imaging. A calibration procedure was developed to compensate for the hysteresis of the magnetic scan coils. A combination of sparse and serpentine scanning routines was tested for a crystalline thin film, gold nanoparticles, and in an in-situ liquid phase STEM experiment. Frame rates of 92, 23 and 5.8 s -1 were achieved for images of a width of 128, 256, and 512 pixels, respectively. The methods described here can be applied to single-particle tracking and analysis of radiation sensitive materials.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:2045-2322
2045-2322
DOI:10.1038/s41598-021-02052-1