Estimation of the Corneal Young’s Modulus In Vivo Based on a Fluid-Filled Spherical-Shell Model with Scheimpflug Imaging

Current intraocular pressure (IOP) measurement using air puff could be erroneous without applying proper corrections. Although noncontact tonometry is not considered to be accurate, it is still popularly used by eye clinics. It is thus necessary to extract the correct information from their results....

Full description

Saved in:
Bibliographic Details
Published inJournal of ophthalmology Vol. 2017; no. 2017; pp. 1 - 11
Main Authors Dai, Chi-An, Shih, Wen-Pin, Wang, I. J., Yen, Jia-Yush, Lin, Hung-Chou, Huang, Tzu-Han, Huang, Chun-Ju, Shih, Po-Jen, Cao, Hui-Jyun
Format Journal Article
LanguageEnglish
Published Cairo, Egypt Hindawi Publishing Corporation 01.01.2017
Hindawi
John Wiley & Sons, Inc
Wiley
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Current intraocular pressure (IOP) measurement using air puff could be erroneous without applying proper corrections. Although noncontact tonometry is not considered to be accurate, it is still popularly used by eye clinics. It is thus necessary to extract the correct information from their results. This study proposes a practical approach to correctly measure IOP in vivo. By embedding a new model-based correction to the Corvis® ST, we can extract the corneal Young’s modulus from the patient data. This Young’s modulus can be used to correct the IOP readings. The tests were applied to 536 right eyes of 536 healthy subjects (228 male and 308 female) between March of 2012 and April of 2016. The tests were applied to patients at the Department of Ophthalmology, National Taiwan University Hospital and the Hung-Chuo Eye Clinics. The statistical analysis showed that the value for the Young’s modulus was independent of all the other parameters collected from the Corvis ST, including the corneal thickness and the intraocular pressure. Therefore, it is important to independently measure the Young’s modulus instead of depending on the correlation with the other parameters. This study adds the methodology of measuring corneal stiffness in vivo for ophthalmologists’ reference in diagnosis.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 14
content type line 23
Academic Editor: Marco Lombardo
ISSN:2090-004X
2090-0058
DOI:10.1155/2017/5410143