快速检测心型脂肪酸结合蛋白的荧光免疫微流体测试卡

研制出一种以时间分辨荧光微球作为标记,自驱动快速检测H-FABP的荧光免疫微流体测试卡。利用激光切割法在双面胶上简便、快速地切割出所设计的微通道结构,并采用激光切割法制作出聚甲基丙烯酸甲酯(PMMA)测试卡底板及上盖。使用提拉涂膜的方法在PMMA底板表面修饰马来酸酐官能团,有效地解决了捕获抗体在PMMA表面的固定问题。使用等离子体处理测试卡上盖改善其亲水性,使液体能够在微通道内自行流动。使用此测试卡可以实现对心型脂肪酸结合蛋白(H-FABP)的快速检测,线性检测范围为0.5~100 ng/mL,检出限为0.1 ng/mL(S/N=3),检测时间少于10 min,批内相对标准偏差(RSD)〈10...

Full description

Saved in:
Bibliographic Details
Published in分析化学 Vol. 44; no. 4; pp. 633 - 639
Main Author 孟凡达 霍卫松 贺美琳 李昊 廉洁 石西增 高云华
Format Journal Article
LanguageChinese
Published 中国科学院理化技术研究所光化学转换和功能材料重点实验室,北京100190 2016
中国科学院大学,北京100049%中国科学院理化技术研究所光化学转换和功能材料重点实验室,北京,100190%中国人民公安大学刑事科学技术系,北京,100038%东莞博识生物科技有限公司,东莞,523808
Subjects
Online AccessGet full text
ISSN0253-3820
DOI10.11895/j.issn.0253-3820.160045

Cover

Loading…
More Information
Summary:研制出一种以时间分辨荧光微球作为标记,自驱动快速检测H-FABP的荧光免疫微流体测试卡。利用激光切割法在双面胶上简便、快速地切割出所设计的微通道结构,并采用激光切割法制作出聚甲基丙烯酸甲酯(PMMA)测试卡底板及上盖。使用提拉涂膜的方法在PMMA底板表面修饰马来酸酐官能团,有效地解决了捕获抗体在PMMA表面的固定问题。使用等离子体处理测试卡上盖改善其亲水性,使液体能够在微通道内自行流动。使用此测试卡可以实现对心型脂肪酸结合蛋白(H-FABP)的快速检测,线性检测范围为0.5~100 ng/mL,检出限为0.1 ng/mL(S/N=3),检测时间少于10 min,批内相对标准偏差(RSD)〈10%,批间RSD〈15%。本方法具有灵敏度高、检测时间短、结果准确等优点,可以满足临床检测的需求,具有良好的应用前景。
Bibliography:Heart-type fatty acid binding protein; Detection; Microfluidic system; Time-resolved fluorescence
22-1125/O6
A kind of easily assembled self-driving test card was developed for the rapid detection and quantification of heart-type fatty acid binding protein( H-FABP) by using time-resolved fluorescence microspheres as signal probes. It was easy to make microfluidic channel structures with double-sided adhesive and cut out of test card substrate and cover based on polymethyl methacrylate( PMMA) material by laser cutting technology. The membrane of polystyrene containing maleic anhydride functional groups was coated on PMMA plate surface by dip-coating method. Further,the capture antibody was immobilized effectively on PMMA surface by covalent binding with maleic anhydride groups. The cover was treated with plasma treatment to improve the hydrophilic ability,so that liquid could flow smoothly in the microchannel. The whole analysis process could be completed within 10 min. There was a very good linear correlation b
ISSN:0253-3820