Determination of tobramycin in eye drops with an open-source hardware ion mobility spectrometer
The analysis of tobramycin was demonstrated successfully as an example for electrospray ionization on an open-source hardware ion mobility spectrometer. This instrument was assembled inexpensively in-house, and required only very few purpose-made components. The quantitative determination of tobramy...
Saved in:
Published in | Analytical and bioanalytical chemistry Vol. 414; no. 14; pp. 4059 - 4066 |
---|---|
Main Authors | , |
Format | Journal Article |
Language | English |
Published |
Berlin/Heidelberg
Springer Berlin Heidelberg
01.06.2022
Springer Springer Nature B.V |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | The analysis of tobramycin was demonstrated successfully as an example for electrospray ionization on an open-source hardware ion mobility spectrometer. This instrument was assembled inexpensively in-house, and required only very few purpose-made components. The quantitative determination of tobramycin required 20 s for a reading. The calibration curve for the range from 50 to 200 μM was found to be linear with a correlation coefficient of
r
= 0.9994. A good reproducibility was obtained (3% relative standard deviation) and the limit of detection was determined as 8 μM. As the concentration of the active ingredient in the eye drops (ophthalmic solutions) is too high for the sensitivity of the instrument, the samples had to be diluted appropriately.
Graphical abstract |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 14 content type line 23 |
ISSN: | 1618-2642 1618-2650 1618-2650 |
DOI: | 10.1007/s00216-022-04050-2 |