Prosthetic overhang is the most effective way to prevent scapular conflict in a reverse total shoulder prosthesis

Background and purpose Despite good clinical results with the reverse total shoulder arthroplasty, inferior scapular notching remains a concern. We evaluated 6 different solutions to overcome the problem of scapular notching. Methods An average and a "worst case scenario" shape in A-P view...

Full description

Saved in:
Bibliographic Details
Published inActa orthopaedica Vol. 81; no. 6; pp. 719 - 726
Main Authors de Wilde, Lieven F, Poncet, Didier, Middernacht, Bart, Ekelund, Anders
Format Journal Article
LanguageEnglish
Published Basingstoke Nordic Orthopaedic Federation 01.12.2010
Taylor & Francis
Informa Healthcare
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Background and purpose Despite good clinical results with the reverse total shoulder arthroplasty, inferior scapular notching remains a concern. We evaluated 6 different solutions to overcome the problem of scapular notching. Methods An average and a "worst case scenario" shape in A-P view in a 2-D computer model of a scapula was created, using data from 200 "normal" scapulae, so that the position of the glenoid and humeral component could be changed as well as design features such as depth of the polyethylene insert, the size of glenosphere, the position of the center of rotation, and downward glenoid inclination. The model calculated the maximum adduction (notch angle) in the scapular plane when the cup of the humeral component was in conflict with the scapula. Results A change in humeral neck shaft inclination from 155° to 145° gave a 10° gain in notch angle. A change in cup depth from 8 mm to 5 mm gave a gain of 12°. With no inferior prosthetic overhang, a lateralization of the center of rotation from 0 mm to 5 mm gained 16°. With an inferior overhang of only 1 mm, no effect of lateralizing the center of rotation was noted. Downward glenoid inclination of 0º to 10º gained 10°. A change in glenosphere radius from 18 mm to 21 mm gained 31° due to the inferior overhang created by the increase in glenosphere. A prosthetic overhang to the bone from 0 mm to 5 mm gained 39°. Interpretation Of all 6 solutions tested, the prosthetic overhang created the biggest gain in notch angle and this should be considered when designing the reverse arthroplasty and defining optimal surgical technique.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:1745-3674
1745-3682
DOI:10.3109/17453674.2010.538354