Liquid chromatography-tandem mass spectrometry analysis of 17-hydroxyprogesterone in dried blood spots revealed matrix effect on immunoassay

Immunoassays for measuring 17-hydroxyprogesterone (17-OHP) produce high rates of false positives that impact the identification of congenital adrenal hyperplasia (CAH) in neonates. A confirmatory test with high analytical specificity employing liquid chromatography-tandem mass spectrometry (LC-MS/MS...

Full description

Saved in:
Bibliographic Details
Published inAnalytical and bioanalytical chemistry Vol. 411; no. 2; pp. 395 - 402
Main Authors Han, Liqiao, Tavakoli, Norma P., Morrissey, Mark, Spink, David C., Cao, Zhimin (Tim)
Format Journal Article
LanguageEnglish
Published Berlin/Heidelberg Springer Berlin Heidelberg 01.01.2019
Springer
Springer Nature B.V
Subjects
Online AccessGet full text
ISSN1618-2642
1618-2650
1618-2650
DOI10.1007/s00216-018-1449-0

Cover

More Information
Summary:Immunoassays for measuring 17-hydroxyprogesterone (17-OHP) produce high rates of false positives that impact the identification of congenital adrenal hyperplasia (CAH) in neonates. A confirmatory test with high analytical specificity employing liquid chromatography-tandem mass spectrometry (LC-MS/MS) methodology is needed in newborn screening for CAH. 17-OHP and cortisol were extracted from dried blood spot (DBS) samples, resolved on a C18 column, and measured using tandem mass spectrometry. The results were compared with those determined using the AutoDELFIA immunoassay. The LC-MS/MS method had a limit of quantitation of 10.0 and 5.0 ng/mL for 17-OHP and cortisol, respectively. The method characteristics showed coefficient variation (%CV) ≤ 11.9% for both 17-OHP and cortisol, recoveries ranging from 83.1 to 101.5% for 17-OHP and from 95.1 to 102.8% for cortisol, and linearity with R 2  = 0.9994 for 17-OHP and R 2  = 0.9996 for cortisol, clinical sensitivity of 100.0% and a specificity of 96.4% as obtained by receiver operating characteristic analysis on 45 patient samples when 17-OHP > 39.1 ng/mL was selected as the cutoff value. Comparison between the LC-MS/MS and the AutoDELFIA immunoassay methods revealed a poor correlation for patient DBS samples ( R 2  = 0.6784); however, an excellent correlation was obtained for QC and proficiency test (PT) DBS samples ( R 2  = 0.9797). The LC-MS/MS method produced reliable results for 17-OHP and cortisol for the diagnosis of CAH. The AutoDELFIA immunoassay appears to be subject to matrix effects in the analysis for 17-OHP in DBS patient samples. The DBS samples of non-patient origin may not be suitable for assessing analytical accuracy of immunoassays.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 14
content type line 23
ISSN:1618-2642
1618-2650
1618-2650
DOI:10.1007/s00216-018-1449-0