Quantification of L-ergothioneine in whole blood by hydrophilic interaction ultra-performance liquid chromatography and UV-detection

A new hydrophilic interaction ultra‐performance LC method was established for the whole blood measurement of L‐ergothioneine. Chromatographic separation was achieved in a fairly short time, less than 4 min, on a 100 × 2.1 mm Acquity UPLC BEH HILIC 1.7 μm column with a mobile phase consisting of a mi...

Full description

Saved in:
Bibliographic Details
Published inJournal of separation science Vol. 36; no. 6; pp. 1002 - 1006
Main Authors Sotgia, Salvatore, Zinellu, Angelo, Pintus, Gianfranco, Pinna, Gerard Aime, Deiana, Luca, Carru, Ciriaco
Format Journal Article
LanguageEnglish
Published Weinheim Blackwell Publishing Ltd 01.03.2013
Wiley
Wiley Subscription Services, Inc
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:A new hydrophilic interaction ultra‐performance LC method was established for the whole blood measurement of L‐ergothioneine. Chromatographic separation was achieved in a fairly short time, less than 4 min, on a 100 × 2.1 mm Acquity UPLC BEH HILIC 1.7 μm column with a mobile phase consisting of a mixture of 100 mmol/L ammonium acetate/ACN/water (5:85:10, v/v/v) that flowed isocratically at 0.250 mL/min. The LOD and the limit of quantification were 3.85 and 11.67 μmol/L, respectively. The method exhibited linearity in a concentration range of 15.63–1000 μmol/L (R2 > 0.999). Mean recovery was 96.34% whereas intraassay and interassay precision were 1.52 and 1.82% RSD, respectively. On the whole, the developed method is simple, fast, precise, accurate, and sensitive and may be useful for routine analyses.
Bibliography:ark:/67375/WNG-598L7XRR-1
istex:F9CAD1E7668AB57DF05067880DC3EB213384877E
ArticleID:JSSC3168
ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ObjectType-Article-2
ObjectType-Feature-1
ISSN:1615-9306
1615-9314
DOI:10.1002/jssc.201201016