Upgrade of the tangentially viewing vacuum ultraviolet (VUV) telescope system for 2D fluctuation measurement in the large helical device

A high-speed tangentially viewing vacuum ultraviolet (VUV) telescope system, using an inverse Schwarzschild-type optic system was developed to study fluctuations in the Large Helical Device (LHD). However, for the original system, the sampling rate was restricted to below 2000 Hz due to the low sign...

Full description

Saved in:
Bibliographic Details
Published inReview of scientific instruments Vol. 87; no. 11; p. 11E307
Main Authors Wang, Z J, Ming, T F, Gao, X, Du, X D, Ohdachi, S
Format Journal Article
LanguageEnglish
Published United States 01.11.2016
Online AccessGet more information

Cover

Loading…
More Information
Summary:A high-speed tangentially viewing vacuum ultraviolet (VUV) telescope system, using an inverse Schwarzschild-type optic system was developed to study fluctuations in the Large Helical Device (LHD). However, for the original system, the sampling rate was restricted to below 2000 Hz due to the low signal to noise (S/N) ratio in the experiment. In order to improve the S/N ratio, upgrade of the system was made. With this upgraded optical system, the maximum framing rate is improved to 6000 fps with a similar spatial resolution. Rotation of the m = 2 structure caused by the magnetohydrodynamic (MHD) instability is measured by the upgraded system. The spatial structure of the image is consistent with the synthetic image assuming the interchange mode type displacement of the flux surfaces.
ISSN:1089-7623
DOI:10.1063/1.4959951