64-slice computed tomography assessment of coronary artery stents: a phantom study
To compare the use of a new 64-slice computed tomography (CT) scanner with 16-slice CT in the visualization of coronary artery stent lumen. Eight different coronary artery stents, each with a diameter of 3 mm, were placed in a static chest phantom. The phantom was positioned in the CT gantry at an a...
Saved in:
Published in | Acta radiologica (1987) Vol. 47; no. 1; p. 36 |
---|---|
Main Authors | , , , , , , , |
Format | Journal Article |
Language | English |
Published |
England
01.02.2006
|
Subjects | |
Online Access | Get more information |
Cover
Loading…
Summary: | To compare the use of a new 64-slice computed tomography (CT) scanner with 16-slice CT in the visualization of coronary artery stent lumen.
Eight different coronary artery stents, each with a diameter of 3 mm, were placed in a static chest phantom. The phantom was positioned in the CT gantry at an angle of 0 degrees and 45 degrees towards the z-axis and examined with both a 64-slice and a 16-slice CT scanner. Effective slice thickness was 0.6 mm with 64-slice CT and 1 mm with 16-slice CT. A reconstruction increment of 0.3 mm was applied in both scanners. Image quality was assessed visually using a 5-point grading scale. Stent diameters were measured and compared using paired Wilcoxon tests.
Artificial lumen reduction was significantly less with 64-slice than with 16-slice CT. Average visible stent lumen was 53.4% using 64-slice CT and 47.5% with 16-slice MSCT. Most severe artifacts were seen in stents with radiopaque markers. Using 64-slice CT, image noise increased by approximately 30% due to thinner slice thickness.
Improved spatial resolution of 64-slice CT resulted in superior assessment of coronary artery stent lumen compared to 16-slice CT. However, a relevant part of the stent lumen is still not assessable with multi-slice CT. |
---|---|