Determination of Lead in Serum by Graphite Furnace Atomic Absorption Spectrometry Using Chromium Nitrate as a Matrix Modifier
A method for the determination of lead in serum by graphite furnace atomic absorption spectrometry with the standard type graphite tube using chromium nitrate as a matrix modifier is described. Serum is used without complicated sample pretreatment but with dilution by 1% Triton X-100 of 0.1 mol/l ni...
Saved in:
Published in | Eisei kagaku Vol. 37; no. 1; pp. 22 - 27 |
---|---|
Main Authors | , , , |
Format | Journal Article |
Language | English |
Published |
Tokyo
The Pharmaceutical Society of Japan
1991
公益社団法人日本薬学会 Nippon Yakugakkai |
Subjects | |
Online Access | Get full text |
ISSN | 0013-273X |
DOI | 10.1248/jhs1956.37.22 |
Cover
Summary: | A method for the determination of lead in serum by graphite furnace atomic absorption spectrometry with the standard type graphite tube using chromium nitrate as a matrix modifier is described. Serum is used without complicated sample pretreatment but with dilution by 1% Triton X-100 of 0.1 mol/l nitric acid solution containing 2000 mgCr/l. Detection limit (S/N=3) was 0.22 μg/l in serum. Reproducibility (RSD) of 4% at n=10 was observed. Recovery of lead in a certified reference material, bovine blood was 104%. The average sensitivity for five different serum samples was 0.00304 abs per μg/l (RSD : 4%). Lead level in 1.3 μg/l of serum was observed in twenty sera of normal subjects. |
---|---|
ISSN: | 0013-273X |
DOI: | 10.1248/jhs1956.37.22 |