Nanodosimetric measurements and calculations in a neutron therapy beam

A comparison of calculated and measured values of the dose mean lineal energy (y(D)) for the former neutron therapy beam at Louvain-la-Neuve is reported. The measurements were made with wall-less tissue-equivalent proportional counters using the variance-covariance method and simulating spheres with...

Full description

Saved in:
Bibliographic Details
Published inRadiation protection dosimetry Vol. 126; no. 1-4; p. 463
Main Authors Grindborg, J-E, Lillhök, J E, Lindborg, L, Gudowska, I, Söderberg, J, Carlsson, G Alm, Nikjoo, H
Format Journal Article
LanguageEnglish
Published England 01.01.2007
Subjects
Online AccessGet more information

Cover

Loading…
More Information
Summary:A comparison of calculated and measured values of the dose mean lineal energy (y(D)) for the former neutron therapy beam at Louvain-la-Neuve is reported. The measurements were made with wall-less tissue-equivalent proportional counters using the variance-covariance method and simulating spheres with diameters between 10 nm and 15 microm. The calculated y(D)-values were obtained from simulated energy distributions of neutrons and charged particles inside an A-150 phantom and from published y(D)-values for mono-energetic ions. The energy distributions of charged particles up to oxygen were determined with the SHIELD-HIT code using an MCNPX simulated neutron spectrum as an input. The mono-energetic ion y(D)-values in the range 3-100 nm were taken from track-structure simulations in water vapour done with PITS/KURBUC. The large influence on the dose mean lineal energy from the light ion (A > 4) absorbed dose fraction, may explain an observed difference between experiment and calculation. The latter being larger than earlier reported result. Below 50 nm, the experimental values increase while the calculated decrease.
ISSN:0144-8420
DOI:10.1093/rpd/ncm093