Aortic sinus flow stasis likely in valve-in-valve transcatheter aortic valve implantation

Abstract Objective Valve-in-valve procedures using transcatheter aortic valves are increasingly performed to treat degenerated bioprosthetic surgical aortic valves because they are less invasive than redo aortic valve replacement. The objective of this study is to quantify the changes in aortic sinu...

Full description

Saved in:
Bibliographic Details
Published inThe Journal of thoracic and cardiovascular surgery Vol. 154; no. 1; pp. 32 - 43.e1
Main Authors Hatoum, Hoda, BS, Moore, Brandon L., PhD, Maureira, Pablo, MD, PhD, Dollery, Jennifer, RN, Crestanello, Juan A., MD, Dasi, Lakshmi Prasad, PhD
Format Journal Article
LanguageEnglish
Published United States Elsevier Inc 01.07.2017
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Abstract Objective Valve-in-valve procedures using transcatheter aortic valves are increasingly performed to treat degenerated bioprosthetic surgical aortic valves because they are less invasive than redo aortic valve replacement. The objective of this study is to quantify the changes in aortic sinus blood flow dynamics before and after a valve-in-valve procedure to gain insight into mechanisms for clinical and subclinical thrombosis of leaflets. Methods A detailed description of the sinus hemodynamics for valve-in-valve implantation was performed in vitro. A Medtronic Hancock II (Medtronic Inc, Minneapolis, Minn) porcine bioprosthesis was modeled as a surgical aortic valve, and Medtronic CoreValve and Edwards Sapien (Edwards Lifesciences, Irvine, Calif) valves were used as the transcatheter aortic valves. High-resolution particle image velocimetry was used to compare the flow patterns from these 2 valves within both the left coronary and noncoronary sinuses in vitro. Results Velocity and vorticity within the surgical valve sinuses reached peak values of 0.7 m/s and 1000 s−1 , with a 70% decrease in peak fluid shear stress near the aortic side of the leaflet in the noncoronary sinus. With the introduction of transcatheter aortic valves, peak velocity and vorticity were reduced to approximately 0.4 m/s and 550 s−1 and 0.58 m/s and 653 s−1 without coronary flow and 0.60 m/s and 631 s−1 and 0.81 m/s and 669 s−1 with coronary flow for the CoreValve and Sapien valve-in-valve implantations, respectively. Peak shear stress was approximately 38% higher along the aortic side of the coronary versus noncoronary transcatheter aortic valve leaflet. Conclusions Decreased flow and shear stress in valve-in-valve procedures indicate a higher risk of leaflet thrombosis secondary to flow stasis, perhaps more so in the noncoronary sinus.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0022-5223
1097-685X
DOI:10.1016/j.jtcvs.2017.03.053