Electromagnetic interference from lasers and intense light sources in the treatment of patients with artificial pacemakers and other implantable cardiac devices

Measurements of the electric and magnetic field strengths surrounding six laser systems and one intense pulsed light system were carried out. The results were compared to exposure limits published by cardiac device manufacturers to assess the risk of electromagnetic interference to implantable cardi...

Full description

Saved in:
Bibliographic Details
Published inLasers in medical science Vol. 30; no. 5; pp. 1619 - 1622
Main Authors Lister, Tom, Grant, Lindsay, Lee, Siu-Man, Cole, Richard P., Jones, Anthony, Taylor, Timothy, Mayo, Angela, Wright, Philip A.
Format Journal Article
LanguageEnglish
Published London Springer London 01.07.2015
Springer Nature B.V
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Measurements of the electric and magnetic field strengths surrounding six laser systems and one intense pulsed light system were carried out. The results were compared to exposure limits published by cardiac device manufacturers to assess the risk of electromagnetic interference to implantable cardiac devices such as pacemakers or implantable cardioverter defibrillators. The majority of lasers assessed in this study were found to produce electric and magnetic field strengths below the published exposure limits for cardiac devices. However, the low-frequency electric field and static magnetic field of both the CO 2 laser and the ruby laser were found to exceed these limits. Ensuring that a small separation is maintained at all times between the laser unit and any patient with a pacemaker or implantable cardioverter defibrillator appears to be a sensible expedient in avoiding overexposure of an implantable cardiac device to electromagnetic interference. Due to the single-shot fast discharge nature of the intense pulsed light system, changes in electromagnetic field strength were too fast for some of the measuring equipment used in this study to register accurate readings during operation.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0268-8921
1435-604X
DOI:10.1007/s10103-013-1470-7