Analyzing the heart rhythm during chest compressions: Performance and clinical value of a new AED algorithm

Automated external defibrillators (AED) prompt the rescuer to stop chest compressions (CC) for ECG analysis during out-of-hospital cardiac arrest (OHCA). We assessed the diagnostic accuracy and clinical benefit of a new AED algorithm (cprINSIGHT), which analyzes ECG and impedance signals during CC,...

Full description

Saved in:
Bibliographic Details
Published inResuscitation Vol. 162; pp. 320 - 328
Main Authors de Graaf, Corina, Beesems, Stefanie G., Oud, Sharon, Stickney, Ronald E., Piraino, Daniel W., Chapman, Fred W., Koster, Rudolph W.
Format Journal Article
LanguageEnglish
Published Ireland Elsevier B.V 01.05.2021
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Automated external defibrillators (AED) prompt the rescuer to stop chest compressions (CC) for ECG analysis during out-of-hospital cardiac arrest (OHCA). We assessed the diagnostic accuracy and clinical benefit of a new AED algorithm (cprINSIGHT), which analyzes ECG and impedance signals during CC, allowing rhythm analysis with ongoing chest compressions. Amsterdam Police and Fire Fighters used a conventional AED in 2016−2017 (control) and an AED with cprINSIGHT in 2018−2019 (intervention). In the intervention AED, cprINSIGHT was activated after the first (conventional) analysis. This algorithm classified the rhythm as “shockable” (S) and “non-shockable” (NS), or “pause needed”. Sensitivity for S, specificity for NS with 90% lower confidence limit (LCL), chest compression fractions (CCF) and pre-shock pause were compared between control and intervention cases accounting for multiple observations per patient. Data from 465 control and 425 intervention cases were analyzed. cprINSIGHT reached a decision during CC in 70% of analyses. Sensitivity of the intervention AED was 96%, (LCL 93%) and specificity was 98% (LCL 97%), both not significantly different from control. Intervention cases had a shorter median pre-shock pause compared to control cases (8 s vs 22 s, p < 0.001) and higher median CCF (86% vs 80%, P < 0.001). AEDs with cprINSIGHT analyzed the ECG during chest compressions in 70% of analyses with 96% sensitivity and 98% specificity when it made a S or a NS decision. Compared to conventional AEDs, cprINSIGHT leads to a significantly shorter pre-shock pause and a significant increase in CCF.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0300-9572
1873-1570
DOI:10.1016/j.resuscitation.2021.01.003