Analyzing the heart rhythm during chest compressions: Performance and clinical value of a new AED algorithm
Automated external defibrillators (AED) prompt the rescuer to stop chest compressions (CC) for ECG analysis during out-of-hospital cardiac arrest (OHCA). We assessed the diagnostic accuracy and clinical benefit of a new AED algorithm (cprINSIGHT), which analyzes ECG and impedance signals during CC,...
Saved in:
Published in | Resuscitation Vol. 162; pp. 320 - 328 |
---|---|
Main Authors | , , , , , , |
Format | Journal Article |
Language | English |
Published |
Ireland
Elsevier B.V
01.05.2021
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Automated external defibrillators (AED) prompt the rescuer to stop chest compressions (CC) for ECG analysis during out-of-hospital cardiac arrest (OHCA). We assessed the diagnostic accuracy and clinical benefit of a new AED algorithm (cprINSIGHT), which analyzes ECG and impedance signals during CC, allowing rhythm analysis with ongoing chest compressions.
Amsterdam Police and Fire Fighters used a conventional AED in 2016−2017 (control) and an AED with cprINSIGHT in 2018−2019 (intervention). In the intervention AED, cprINSIGHT was activated after the first (conventional) analysis. This algorithm classified the rhythm as “shockable” (S) and “non-shockable” (NS), or “pause needed”. Sensitivity for S, specificity for NS with 90% lower confidence limit (LCL), chest compression fractions (CCF) and pre-shock pause were compared between control and intervention cases accounting for multiple observations per patient.
Data from 465 control and 425 intervention cases were analyzed. cprINSIGHT reached a decision during CC in 70% of analyses. Sensitivity of the intervention AED was 96%, (LCL 93%) and specificity was 98% (LCL 97%), both not significantly different from control. Intervention cases had a shorter median pre-shock pause compared to control cases (8 s vs 22 s, p < 0.001) and higher median CCF (86% vs 80%, P < 0.001).
AEDs with cprINSIGHT analyzed the ECG during chest compressions in 70% of analyses with 96% sensitivity and 98% specificity when it made a S or a NS decision. Compared to conventional AEDs, cprINSIGHT leads to a significantly shorter pre-shock pause and a significant increase in CCF. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 0300-9572 1873-1570 |
DOI: | 10.1016/j.resuscitation.2021.01.003 |