The 'virtual DBS population': five realistic computational models of deep brain stimulation patients for electromagnetic MR safety studies

We design, develop, and disseminate a 'virtual population' of five realistic computational models of deep brain stimulation (DBS) patients for electromagnetic (EM) analysis. We found five DBS patients in our institution' research patient database who received high quality post-DBS sur...

Full description

Saved in:
Bibliographic Details
Published inPhysics in medicine & biology Vol. 64; no. 3; p. 35021
Main Authors Guerin, Bastien, Iacono, Maria Ida, Davids, Mathias, Dougherty, Darin, Angelone, Leonardo M, Wald, Lawrence L
Format Journal Article
LanguageEnglish
Published England IOP Publishing 04.02.2019
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:We design, develop, and disseminate a 'virtual population' of five realistic computational models of deep brain stimulation (DBS) patients for electromagnetic (EM) analysis. We found five DBS patients in our institution' research patient database who received high quality post-DBS surgery computer tomography (CT) examinations of the head and neck. Three patients have a single implanted pulse generator (IPG) and the two others have two IPGs (one for each lead). Moreover, one patient has two abandoned leads on each side of the head. For each patient, we combined the head and neck volumes into a 'virtual CT', from which we extracted the full-length DBS path including the IPG, extension cables, and leads. We corrected topology errors in this path, such as self-intersections, using a previously published optimization procedure. We segmented the virtual CT volume into bones, internal air, and soft tissue classes and created two-manifold, watertight surface meshes of these distributions. In addition, we added a segmented model of the brain (grey matter, white matter, eyes and cerebrospinal fluid) to one of the model (nickname Freddie) that was derived from a T1-weighted MR image obtained prior to the DBS implantation. We simulated the EM fields and specific absorption rate (SAR) induced at 3 Tesla by a quadrature birdcage body coil in each of the five patient models using a co-simulation strategy. We found that inter-subject peak SAR variability across models was independent of the target averaging mass and equal to ~45%. In our simulations of the full brain segmentation and six simplified versions of the Freddie model, the error associated with incorrect dielectric property assignment around the DBS electrodes was greater than the error associated with modeling the whole model as a single tissue class. Our DBS patient models are freely available on our lab website (Webpage of the Martinos Center Phantom Resource 2018 https://phantoms.martinos.org/Main_Page).
Bibliography:PMB-107743.R1
Institute of Physics and Engineering in Medicine
ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0031-9155
1361-6560
1361-6560
DOI:10.1088/1361-6560/aafce8