Determination of uranium and thorium in high purity titanium by ICP-MS after ion exchange separation
The determination of trace amounts of U and Th in Ti by ICP-MS has been investigated. Sample was decomposed with hydrochloric acid in a Teflon pressure vessel, and subsequently added hydrogen peroxide solution. In order to reject matrix effect in the measurement, U was separated from Ti by anion exc...
Saved in:
Published in | BUNSEKI KAGAKU Vol. 42; no. 1; pp. T19 - T22 |
---|---|
Main Authors | , , , |
Format | Journal Article |
Language | English Japanese |
Published |
Tokyo
The Japan Society for Analytical Chemistry
01.01.1993
Nippon Bunseki Kagakkai Japan Science and Technology Agency |
Subjects | |
Online Access | Get full text |
ISSN | 0525-1931 |
DOI | 10.2116/bunsekikagaku.42.T19 |
Cover
Summary: | The determination of trace amounts of U and Th in Ti by ICP-MS has been investigated. Sample was decomposed with hydrochloric acid in a Teflon pressure vessel, and subsequently added hydrogen peroxide solution. In order to reject matrix effect in the measurement, U was separated from Ti by anion exchange using 9 M hydrochloric acid and Th was separated from Ti by cation exchange using 1.5 M hydrochloric acid. Quantitative recoveries were obtained for U and Th. The present method effectively determined U and Th of sub ppb level in Ti. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 14 |
ISSN: | 0525-1931 |
DOI: | 10.2116/bunsekikagaku.42.T19 |