Recalibration of the point-of-care test for CARDIAC T Quantitative with Elecsys Troponin T 3rd generation

The rapid troponin T assay CARDIAC T Quantitative was recalibrated using Elecsys Troponin T 3rd Generation as a new reference method. This paper presents the method comparisons at six centres using the new reference method. Method comparison between CARDIAC T Quantitative versus Elecsys Troponin T 3...

Full description

Saved in:
Bibliographic Details
Published inClinica chimica acta Vol. 307; no. 1-2; pp. 197 - 203
Main Authors Collinson, Paul O, Jørgensen, Bo, Sylvén, Christer, Haass, Markus, Chwallek, Frank, Katus, Hugo A, Müller-Bardorff, Margit, Derhaschnig, Ulla, Hirschl, Michael M, Zerback, Rainer
Format Journal Article Conference Proceeding
LanguageEnglish
Published Shannon Elsevier B.V 01.05.2001
Elsevier
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:The rapid troponin T assay CARDIAC T Quantitative was recalibrated using Elecsys Troponin T 3rd Generation as a new reference method. This paper presents the method comparisons at six centres using the new reference method. Method comparison between CARDIAC T Quantitative versus Elecsys Troponin T 3rd Generation were performed using 319 samples from patients with acute coronary syndromes. The quality of the CARDIAC T Quantitative was controlled by a daily single determination of CARDIAC Control Troponin T, and for the Elecsys Troponin T 3rd Generation, the Elecsys controls were included in each run. The results for the control materials for the CARDIAC T Quantitative were between 93% and 107% of the target values. The CV ranged from 7% to 16%. From the regression analysis, according to Bablok and Passing (y=1.07x) and the Bland and Altman plot, the bias between CARDIAC T Quantitative and Elecsys Troponin T 3rd Generation is from +6% to +7%. The correlation coefficient is 0.93, and a 3×3 comparison of the clinical efficiency yielded 92% clinical concordance between CARDIAC T Quantitative and Elecsys Troponin T 3rd Generation. In conclusion, CARDIAC T Quantitative was in good agreement with the reference and calibration method Elecsys Troponin T 3rd Generation.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0009-8981
1873-3492