Potentiometric Flow Injection Analysis of Glucose Using Hexacyanoferrate(III)-Hexacyanoferrate(II) Potential Buffer
Flow injection analysis of glucose is described, using a combination of the potentiometric detection with the potential buffer solution consisting of hexacyanoferrate(III)-hexacyanoferrate(II). Effects of the concentration of the potential buffer solution, reaction time and temperature are examined...
Saved in:
Published in | Analytical Sciences Vol. 3; no. 5; pp. 453 - 456 |
---|---|
Main Authors | , , , |
Format | Journal Article |
Language | English |
Published |
Tokyo
The Japan Society for Analytical Chemistry
1987
Japan Society for Analytical Chemistry Japan Science and Technology Agency |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Flow injection analysis of glucose is described, using a combination of the potentiometric detection with the potential buffer solution consisting of hexacyanoferrate(III)-hexacyanoferrate(II). Effects of the concentration of the potential buffer solution, reaction time and temperature are examined on sensitivity of the proposed analytical method. Glucose from 10-7M to 10-3M, for injections of 140μl samples, can be determined by changing concentrations of the buffer solution from 1×10-5M to 1×10-2M. |
---|---|
ISSN: | 0910-6340 1348-2246 |
DOI: | 10.2116/analsci.3.453 |