Crown-Cut Endobronchial Ultrasound Guided Transbronchial Aspiration Needle: First Real-World Experiences
Advancements in personalized medicine have increased the demand for quantity and preservation of tissue architecture of endobronchial ultrasound-guided transbronchial needle aspiration (EBUS-TBNA) samples. These demands may be addressed by the SonoTip TopGain needle, which has a 3-point crown-cut de...
Saved in:
Published in | Journal of clinical medicine Vol. 11; no. 1; p. 163 |
---|---|
Main Authors | , , , , , , |
Format | Journal Article |
Language | English |
Published |
Switzerland
MDPI AG
29.12.2021
MDPI |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Advancements in personalized medicine have increased the demand for quantity and preservation of tissue architecture of endobronchial ultrasound-guided transbronchial needle aspiration (EBUS-TBNA) samples. These demands may be addressed by the SonoTip TopGain
needle, which has a 3-point crown-cut design that contrasts with the standard single bevel design of the ViziShot 2
. The objective was to compare the SonoTip TopGain
and ViziShot 2
needles by considering biopsy sample characteristics, diagnostic accuracy, and patient safety. The primary endpoint of the study was the number of high-power fields (HPFs) in the center of the formalin-fixed paraffin-embedded cell block per sample. The lymph node with the highest probability for malignant infiltration based on size and sonographic appearance was chosen as the target lymph node for 20 patients. The same lymph node in each patient was sampled using both the ViziShot 2
and SonoTip TopGain
needles. The samples were measured, sliced, and analyzed by a pathologist. Sixteen patients were biopsied with both needles. Four patients could not be biopsied with the SonoTip TopGain
needle since it could not penetrate cartilage or be repositioned to bypass cartilage. HPFs and sample dimensions were significantly greater in the patients where sampling with the SonoTip TopGain
needle was possible (
= 0.007 and
= 0.005, respectively). Diagnostic accuracy and safety profiles were comparable. Significantly more material can be sampled using the SonoTip TopGain
needle when cartilage penetration can be avoided. This improves the yield for molecular workup in the era of personalized medicine. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 These authors contributed equally to this work. |
ISSN: | 2077-0383 2077-0383 |
DOI: | 10.3390/jcm11010163 |