Determination of Zinc, Cadmium, Lead, Copper and Silver Using a Carbon Paste Electrode and a Screen Printed Electrode Modified with Chromium(III) Oxide

In this study, the preparation and electrochemical application of a chromium(III) oxide modified carbon paste electrode (Cr-CPE) and a screen printed electrode (SPE), made from the same material and optimized for the simple, cheap and sensitive simultaneous determination of zinc, cadmium, lead, copp...

Full description

Saved in:
Bibliographic Details
Published inSensors (Basel, Switzerland) Vol. 17; no. 8; p. 1832
Main Authors Koudelkova, Zuzana, Syrovy, Tomas, Ambrozova, Pavlina, Moravec, Zdenek, Kubac, Lubomir, Hynek, David, Richtera, Lukas, Adam, Vojtech
Format Journal Article
LanguageEnglish
Published Switzerland MDPI AG 09.08.2017
MDPI
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:In this study, the preparation and electrochemical application of a chromium(III) oxide modified carbon paste electrode (Cr-CPE) and a screen printed electrode (SPE), made from the same material and optimized for the simple, cheap and sensitive simultaneous determination of zinc, cadmium, lead, copper and the detection of silver ions, is described. The limits of detection and quantification were 25 and 80 µg·L for Zn(II), 3 and 10 µg·L for Cd(II), 3 and 10 µg·L for Pb(II), 3 and 10 µg·L for Cu(II), and 3 and 10 µg·L for Ag(I), respectively. Furthermore, this promising modification was transferred to the screen-printed electrode. The limits of detection for the simultaneous determination of zinc, cadmium, copper and lead on the screen printed electrodes were found to be 350 µg·L for Zn(II), 25 µg·L for Cd(II), 3 µg·L for Pb(II) and 3 µg·L for Cu(II). Practical usability for the simultaneous detection of these heavy metal ions by the Cr-CPE was also demonstrated in the analyses of wastewaters.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:1424-8220
1424-8220
DOI:10.3390/s17081832