Operator Radiation and the Efficacy of Ceiling-Suspended Lead Screen Shielding during Coronary Angiography: An Anthropomorphic Phantom Study Using Real-Time Dosimeters

Operator radiation and the radiation protection efficacy of a ceiling-suspended lead screen were assessed during coronary angiography (CA) in a catheterization laboratory. An anthropomorphic phantom was placed under the X-ray beam to simulate patient attenuation in eight CA projections. Using real-t...

Full description

Saved in:
Bibliographic Details
Published inScientific reports Vol. 7; no. 1; p. 42077
Main Authors Jia, Qianjun, Chen, Ziman, Jiang, Xianxian, Zhao, Zhenjun, Huang, Meiping, Li, Jiahua, Zhuang, Jian, Liu, Xiaoqing, Hu, Tianyu, Liang, Wensheng
Format Journal Article
LanguageEnglish
Published England Nature Publishing Group 07.02.2017
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Operator radiation and the radiation protection efficacy of a ceiling-suspended lead screen were assessed during coronary angiography (CA) in a catheterization laboratory. An anthropomorphic phantom was placed under the X-ray beam to simulate patient attenuation in eight CA projections. Using real-time dosimeters, radiation dose rates were measured on models mimicking a primary operator (PO) and an assistant. Subsequently, a ceiling-suspended lead screen was placed in three commonly used positions to compare the radiation protection efficacy. The radiation exposure to the PO was 2.3 to 227.9 (mean: 67.2 ± 49.0) μSv/min, with the left anterior oblique (LAO) 45°/cranial 25° and cranial 25° projections causing the highest and the lowest dose rates, respectively. The assistant experienced significantly less radiation overall (mean: 20.1 ± 19.6 μSv/min, P < 0.003), with the right anterior oblique (RAO) 30° and cranial 25° projections resulting in the highest and lowest exposure levels, respectively. Combined with table-side shielding, the ceiling-suspended lead screen reduced the radiation to the PO by 76.8%, 81.9% and 93.5% when placed close to the patient phantom, at the left side and close to the PO, respectively, and reduced the radiation to the assistant by 70.3%, 76.7% and 90.0%, respectively. When placed close to the PO, a ceiling-suspended lead screen provides substantial radiation protection during CA.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
These authors contributed equally to this work.
ISSN:2045-2322
2045-2322
DOI:10.1038/srep42077