False ventricular tachycardia alarm suppression in the ICU based on the discrete wavelet transform in the ECG signal

Abstract Over the past few years, reducing the number of false positive cardiac monitor alarms (FA) in the intensive care unit (ICU) has become an issue of the utmost importance. In our work, we developed a robust methodology that, without the need for additional non-ECG waveforms, suppresses false...

Full description

Saved in:
Bibliographic Details
Published inJournal of electrocardiology Vol. 47; no. 6; pp. 775 - 780
Main Authors Salas-Boni, Rebeca, PhD, Bai, Yong, MS, Harris, Patricia Rae Eileen, PhD, RN, Drew, Barbara J., RN, PhD, FAAN, FAHA, Hu, Xiao, PhD
Format Journal Article
LanguageEnglish
Published United States Elsevier Inc 01.11.2014
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Abstract Over the past few years, reducing the number of false positive cardiac monitor alarms (FA) in the intensive care unit (ICU) has become an issue of the utmost importance. In our work, we developed a robust methodology that, without the need for additional non-ECG waveforms, suppresses false positive ventricular tachycardia (VT) alarms without resulting in false negative alarms. Our approach is based on features extracted from the ECG signal 20 seconds prior to a triggered alarm. We applied a multi resolution wavelet transform to the ECG data 20 seconds prior to the alarm trigger, extracted features from appropriately chosen scales and combined them across all available leads. These representations are presented to a L1-regularized logistic regression classifier. Results are shown in two datasets of physiological waveforms with manually assessed cardiac monitor alarms: the MIMIC II dataset, where we achieved a false alarm (FA) suppression of 21% with zero true alarm (TA) suppression; and a dataset compiled by UCSF and General Electric, where a 36% FA suppression was achieved with a zero TA suppression. The methodology described in this work could be implemented to reduce the number of false monitor alarms in other arrhythmias.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0022-0736
1532-8430
DOI:10.1016/j.jelectrocard.2014.07.016