Optimal fast T2-weighted magnetic resonance microscopy imaging of the eye and its clinical application
Purpose: To compare a half‐Fourier single‐shot rapid acquisition with relaxation enhancement (RARE) sequence with a balanced steady‐state free precession (b‐SSFP) sequence in the evaluation of the eye using magnetic resonance (MR) microscopy imaging and to clarify the usefulness of RARE microscopy i...
Saved in:
Published in | Journal of magnetic resonance imaging Vol. 31; no. 5; pp. 1210 - 1214 |
---|---|
Main Authors | , , , |
Format | Journal Article |
Language | English |
Published |
Hoboken
Wiley Subscription Services, Inc., A Wiley Company
01.05.2010
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Purpose:
To compare a half‐Fourier single‐shot rapid acquisition with relaxation enhancement (RARE) sequence with a balanced steady‐state free precession (b‐SSFP) sequence in the evaluation of the eye using magnetic resonance (MR) microscopy imaging and to clarify the usefulness of RARE microscopy imaging in evaluating nonoperative glaucoma patients and patients who have undergone surgery for glaucoma or cataract.
Materials and Methods:
One‐mm and 2‐mm slice thickness images of RARE sequence and b‐SSFP sequence using a 1.5 T MR unit and a 23‐mm microscopy coil were obtained in eight healthy volunteers. The signal‐to‐noise (S/N) ratio of aqueous humor in the anterior chamber was measured quantitatively and visualization of the anterior chamber anatomy was assessed qualitatively. Furthermore, we evaluated 21 glaucoma patients (including six postoperative patients) and four patients after cataract surgery with 2‐mm slice thickness RARE MRI.
Results:
The 2‐mm slice thickness RARE imaging had a significantly greater S/N ratio than the 1‐mm slice thickness RARE imaging (P < 0.05) and acquired the best image quality among the four types of images (P < 0.01). Additionally, 2‐mm slice thickness RARE microscopy imaging could depict anterior chamber anatomy of glaucoma eyes and eyes after cataract surgery.
Conclusion:
We believe that optimal fast T2‐weighted MR microimaging might become a useful ophthalmologic examination technique. J. Magn. Reson. Imaging 2010;31:1210–1214. ©2010 Wiley‐Liss, Inc. |
---|---|
Bibliography: | ArticleID:JMRI22165 istex:EF70C928BC4D7FB70902866965C9EF7E36A8CC6D ark:/67375/WNG-42VP07NL-6 ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 1053-1807 1522-2586 |