Lead explantation late after atrial perforation

This report describes the case of a patient in whom atrial perforation with penetration of the thoracic wall was diagnosed 2 years after the implantation of an Accufix lead. Despite this complication, atrial detection in the bipolar mode and ventricular pacing were normal. Digital fluoroscopy detect...

Full description

Saved in:
Bibliographic Details
Published inPacing and clinical electrophysiology Vol. 19; no. 8; p. 1268
Main Authors Trigano, A J, Caus, T
Format Journal Article
LanguageEnglish
Published United States 01.08.1996
Subjects
Online AccessGet more information

Cover

Loading…
More Information
Summary:This report describes the case of a patient in whom atrial perforation with penetration of the thoracic wall was diagnosed 2 years after the implantation of an Accufix lead. Despite this complication, atrial detection in the bipolar mode and ventricular pacing were normal. Digital fluoroscopy detected a fracture with extrusion of a short segment of the retention wire. The rupture of the retention wire might have been the result, but was not the cause of the perforation.
ISSN:0147-8389