High grade glioma radiation therapy on a high field 1.5 Tesla MR-Linac - workflow and initial experience with daily adapt-to-position (ATP) MR guidance: A first report

This study reports the workflow and initial clinical experience of high grade glioma (HGG) radiotherapy on the 1.5 T MR-Linac (MRL), with a focus on the temporal variations of the tumor and feasibility of multi-parametric image (mpMRI) acquisition during routine treatment workflow. Ten HGG patients...

Full description

Saved in:
Bibliographic Details
Published inFrontiers in oncology Vol. 12; p. 1060098
Main Authors Tseng, Chia-Lin, Chen, Hanbo, Stewart, James, Lau, Angus Z, Chan, Rachel W, Lawrence, Liam S P, Myrehaug, Sten, Soliman, Hany, Detsky, Jay, Lim-Fat, Mary Jane, Lipsman, Nir, Das, Sunit, Heyn, Chinthaka, Maralani, Pejman J, Binda, Shawn, Perry, James, Keller, Brian, Stanisz, Greg J, Ruschin, Mark, Sahgal, Arjun
Format Journal Article
LanguageEnglish
Published Switzerland Frontiers Media S.A 28.11.2022
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:This study reports the workflow and initial clinical experience of high grade glioma (HGG) radiotherapy on the 1.5 T MR-Linac (MRL), with a focus on the temporal variations of the tumor and feasibility of multi-parametric image (mpMRI) acquisition during routine treatment workflow. Ten HGG patients treated with radiation within the first year of the MRL's clinical operation, between October 2019 and August 2020, were identified from a prospective database. Workflow timings were recorded and online adaptive plans were generated using the Adapt-To-Position (ATP) workflow. Temporal variation within the FLAIR hyperintense region (FHR) was assessed by the relative FHR volumes (n = 281 contours) and migration distances (maximum linear displacement of the volume). Research mpMRIs were acquired on the MRL during radiation and changes in selected functional parameters were investigated within the FHR. All patients completed radiotherapy to a median dose of 60 Gy (range, 54-60 Gy) in 30 fractions (range, 30-33), receiving a total of 287 fractions on the MRL. The mean in-room time per fraction with or without post-beam research imaging was 42.9 minutes (range, 25.0-69.0 minutes) and 37.3 minutes (range, 24.0-51.0 minutes), respectively. Three patients (30%) required re-planning between fractions 9 to 12 due to progression of tumor and/or edema identified on daily MRL imaging. At the 10, 20, and 30-day post-first fraction time points 3, 3, and 4 patients, respectively, had a FHR volume that changed by at least 20% relative to the first fraction. Research mpMRIs were successfully acquired on the MRL. The median apparent diffusion coefficient (ADC) within the FHR and the volumes of FLAIR were significantly correlated when data from all patients and time points were pooled ( =0.68, <.001). We report the first clinical series of HGG patients treated with radiotherapy on the MRL. The ATP workflow and treatment times were clinically acceptable, and daily online MRL imaging triggered adaptive re-planning for selected patients. Acquisition of mpMRIs was feasible on the MRL during routine treatment workflow. Prospective clinical outcomes data is anticipated from the ongoing UNITED phase 2 trial to further refine the role of MR-guided adaptive radiotherapy.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
This article was submitted to Radiation Oncology, a section of the journal Frontiers in Oncology
Edited by: Enis Ozyar, Acıbadem University, Turkey
Reviewed by: Archya Dasgupta, Tata Memorial Hospital, India; Colette Shen, University of North Carolina Hospitals, United States
ISSN:2234-943X
2234-943X
DOI:10.3389/fonc.2022.1060098