Analysis of image defects in digital intraoral radiography based on photostimulable phosphor plates

Objectives In digital intraoral radiography (DIR), images with defects caused by the digital process are sometimes produced. Hence, we analyzed DIR images with defects taken using the photostimulable phosphor (PSP) plate method and then classified these images based on the causes of the defect. The...

Full description

Saved in:
Bibliographic Details
Published inOral radiology Vol. 39; no. 2; pp. 355 - 363
Main Authors Tashiro, Masayasu, Nakatani, Atsutoshi, Sugiura, Kazutaka, Nakayama, Eiji
Format Journal Article
LanguageEnglish
Published Singapore Springer Nature Singapore 01.04.2023
Springer Nature B.V
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Objectives In digital intraoral radiography (DIR), images with defects caused by the digital process are sometimes produced. Hence, we analyzed DIR images with defects taken using the photostimulable phosphor (PSP) plate method and then classified these images based on the causes of the defect. The aim of this study was to classify defects in DIR using the PSP plate system, and to discuss the causes of each type of defect and the methods that can prevent their occurrence. Methods Images with defects due to the two PSP plate systems (Arcana and Arcana Mira) were selected and classified according to the defect. Image defects caused by geometrical techniques that occur in both the film and PSP plate methods were excluded from this study. Moreover, changes in the rate of occurrence of images with defects over time were analyzed in this study. Results The defects in images obtained by the PSP plate system were classified into six types, which were further divided into those caused by machine error or human error. Both types of error could influence the diagnostic performance. Machine error accidentally and rarely occurred; thus, the prevention of such errors is difficult. However, human error, especially errors caused by photo-induced discharge, could be prevented using appropriate measures. Conclusions In DIR systems using the PSP plate method, human error should be prevented by education and training to improve operation.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0911-6028
1613-9674
DOI:10.1007/s11282-022-00645-8