Electrochemical Detector for High-Performance Liquid Chromatography. VI. Application of a Twin Electrochemical Detector
The fundamental characteristics of a twin electrochemical detector were evaluated to investigate its suitability as a detector for high-performance liquid chromatography. This novel detector was proved to have the following advantages compared to the conventional three-electrode detector. 1) Simulta...
Saved in:
Published in | Chemical & pharmaceutical bulletin Vol. 31; no. 10; pp. 3571 - 3578 |
---|---|
Main Authors | , , , , , , , , |
Format | Journal Article |
Language | English |
Published |
Tokyo
The Pharmaceutical Society of Japan
1983
Maruzen Japan Science and Technology Agency |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | The fundamental characteristics of a twin electrochemical detector were evaluated to investigate its suitability as a detector for high-performance liquid chromatography. This novel detector was proved to have the following advantages compared to the conventional three-electrode detector. 1) Simultaneous determination of reduced and oxidized species is possible, 2) the stability of the electrode is improved by applying a larger potential to the other electrode, 3) the specificity of detection is improved by the re-electrolysis at the second electrode of the reaction product produced at the first electrode, and 4) specific detection of the post-discharging materials and sensitive detection are possible by using appropriate operating circuits. |
---|---|
ISSN: | 0009-2363 1347-5223 |
DOI: | 10.1248/cpb.31.3571 |