Restoration of the physiological flexor/extensor mechanism at the knee in terms of appropriate muscular lever arms, proper required quadriceps force, and suitable patellofemoral compressive force, is fundamental for the success of total knee replacement. Therefore, measurements of anteroposterior tr...

Full description

Saved in:
Bibliographic Details
Published inJournal of orthopaedic research Vol. 24; no. 12; pp. 2222 - 2229
Main Authors Fantozzi, Silvia, Catani, Fabio, Ensini, Andrea, Leardini, Alberto, Giannini, Sandro
Format Journal Article
LanguageEnglish
Published Hoboken Wiley Subscription Services, Inc., A Wiley Company 01.12.2006
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Restoration of the physiological flexor/extensor mechanism at the knee in terms of appropriate muscular lever arms, proper required quadriceps force, and suitable patellofemoral compressive force, is fundamental for the success of total knee replacement. Therefore, measurements of anteroposterior translation of the femoral component over the tibial base‐plate against joint flexion during daily living activities are essential for the assessment of the in vivo performance of current prosthesis designs. Patients treated with posterior stabilized and cruciate retaining prostheses with excellent clinical scores were evaluated during stair climbing, sitting and rising from a chair, and step up and down, using a three‐dimensional pose reconstruction technique based on videofluoroscopy. The posterior stabilized patients experienced a fairly consistent and physiological rollback specific of each motor task, demonstrating proper function of the spine‐cam mechanism. Rollback was somehow inconsistent among subjects in the cruciate retaining group, accompanied with a smaller range of knee flexion. In this group, more posterior locations of the condyles correlated significantly with higher clinical and functional scores. Articular surface conformity restores physiological rollback in the presence of a spine‐cam mechanism, but not coherently in the presence of the posterior cruciate ligament. Published by Wiley Periodicals, Inc. J Orthop Res
Bibliography:ArticleID:JOR20306
istex:2A0B30FD1913AB217A2E1064C525161BD2BC556B
ark:/67375/WNG-0V8BKV6Q-C
ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0736-0266
1554-527X
DOI:10.1002/jor.20306