Improved SENSE imaging using accurate coil sensitivity maps generated by a global magnitude-phase fitting method

Purpose To develop a novel coil sensitivity processing technique that is able to reduce or eliminate aliasing artifacts and noise amplification in image‐domain parallel imaging (i.e., SENSE). Methods Initial coil sensitivity maps were derived from the center k‐space lines by a common self‐calibratio...

Full description

Saved in:
Bibliographic Details
Published inMagnetic resonance in medicine Vol. 74; no. 1; pp. 217 - 224
Main Authors Ma, Ya-Jun, Liu, Wentao, Tang, Xin, Gao, Jia-Hong
Format Journal Article
LanguageEnglish
Published United States Blackwell Publishing Ltd 01.07.2015
Wiley Subscription Services, Inc
Subjects
Online AccessGet full text
ISSN0740-3194
1522-2594
1522-2594
DOI10.1002/mrm.25375

Cover

Loading…
More Information
Summary:Purpose To develop a novel coil sensitivity processing technique that is able to reduce or eliminate aliasing artifacts and noise amplification in image‐domain parallel imaging (i.e., SENSE). Methods Initial coil sensitivity maps were derived from the center k‐space lines by a common self‐calibration method. Then continuous trigonometric functions were used to fit both magnitude and phase maps of the self‐calibrated coil sensitivity profile globally. Further, the global fitted coil sensitivity maps were adopted for SENSE reconstruction. Numerical simulations, as well as experiments on phantoms and human subjects were performed to evaluate and compare the effectiveness of this global magnitude‐phase fitting approach with traditional local fitting methods. Results Both simulation and experimental results demonstrated that the proposed novel global fitting method was able to obtain accurate coil sensitivity profiles without Gibbs oscillations. The resultant SENSE images were improved substantially in terms of aliasing imaging artifacts. Conclusion A global magnitude‐phase fitting method for better estimation of accurate coil sensitivity maps was developed, and it was successfully used in producing high‐quality parallel images. Magn Reson Med 74:217–224, 2015. © 2014 Wiley Periodicals, Inc.
Bibliography:istex:CFB0EF54556B339A18B68DC118C16EE05FFCEA21
ark:/67375/WNG-ZWX9C4ZC-3
Natural Science Foundation of China - No. 81227003
ArticleID:MRM25375
China's National Strategic Basic Research Program - No. (973) (2012CB720700
ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 14
content type line 23
ISSN:0740-3194
1522-2594
1522-2594
DOI:10.1002/mrm.25375