Optimization of pulse sequences in ultrafast magnetic resonance imaging for the diagnosis of acute abdominal pain caused by gastrointestinal disease
Background Magnetic resonance imaging (MRI) is widely used to diagnose acute abdominal pain; however, it remains unclear which pulse sequence has priority in acute abdominal pain. Purpose To investigate the diagnostic accuracy of MRI and to assess the conspicuity of each pulse sequence for the diagn...
Saved in:
Published in | Acta radiologica open Vol. 9; no. 8; p. 2058460120949246 |
---|---|
Main Authors | , , , , , |
Format | Journal Article |
Language | English |
Published |
London, England
SAGE Publications
01.08.2020
Sage Publications Ltd |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Background
Magnetic resonance imaging (MRI) is widely used to diagnose acute abdominal pain; however, it remains unclear which pulse sequence has priority in acute abdominal pain.
Purpose
To investigate the diagnostic accuracy of MRI and to assess the conspicuity of each pulse sequence for the diagnosis of acute abdominal pain due to gastrointestinal diseases
Material and Methods
We retrospectively enrolled 60 patients with acute abdominal pain who underwent MRI for axial and coronal T2-weighted (T2W) imaging, fat-suppressed (FS)-T2W imaging, and true-fast imaging with steady-state precession (True-FISP) and axial T1-weighted (T1W) imaging and investigated the diagnosis with endoscopy, surgery, histopathology, computed tomography, and clinical follow-up as standard references. Two radiologists determined the diagnosis with MRI and rated scores of the respective sequences in assessing intraluminal, intramural, and extramural abnormality using a 5-point scale after one month. Diagnostic accuracy was calculated and scores were compared by Wilcoxon-signed rank test with Bonferroni correction.
Results
Diagnostic accuracy was 90.0% and 93.3% for readers 1 and 2, respectively. Regarding intraluminal abnormality, T2W, FS-T2W, and True-FISP imaging were superior to T1W imaging in both readers. FS-T2W imaging was superior to True-FISP in reader 2 (P < 0.0083). For intramural findings, there was no significant difference in reader 1, whereas T2W, FS-T2W, and True-FISP imaging were superior to T1W imaging in reader 2 (P < 0.0083). For extramural findings, FS-T2W imaging was superior to T2W, T1W, and True-FISP imaging in both readers (P < 0.0083).
Conclusion
T2W and FS-T2W imaging are pivotal pulse sequences and should be obtained before T1W and True-FISP imaging. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 14 content type line 23 |
ISSN: | 2058-4601 2058-4601 |
DOI: | 10.1177/2058460120949246 |