Nontraumatic tibial polyethylene insert cone fracture in mobile-bearing posterior-stabilized total knee arthroplasty

Abstract A 72-year-old male patient underwent mobile-bearing posterior-stabilized total knee arthroplasty for osteoarthritis. He experienced a nontraumatic polyethylene tibial insert cone fracture 27 months after surgery. Scanning electron microscopy of the fracture surface of the tibial insert cone...

Full description

Saved in:
Bibliographic Details
Published inArthroplasty today Vol. 2; no. 4; pp. 157 - 163
Main Authors Tanikake, Yohei, Hayashi, Koji, Ogawa, Munehiro, Inagaki, Yusuke, Kawate, Kenji, Tomita, Tetsuya, Tanaka, Yasuhito
Format Journal Article
LanguageEnglish
Published United States Elsevier Inc 01.12.2016
Elsevier
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Abstract A 72-year-old male patient underwent mobile-bearing posterior-stabilized total knee arthroplasty for osteoarthritis. He experienced a nontraumatic polyethylene tibial insert cone fracture 27 months after surgery. Scanning electron microscopy of the fracture surface of the tibial insert cone suggested progress of ductile breaking from the posterior toward the anterior of the cone due to repeated longitudinal bending stress, leading to fatigue breaking at the anterior side of the cone, followed by the tibial insert cone fracture at the anterior side of the cone, resulting in fracture at the base of the cone. This analysis shows the risk of tibial insert cone fracture due to longitudinal stress in mobile-bearing posterior-stabilized total knee arthroplasty in which an insert is designed to highly conform to the femoral component.
ISSN:2352-3441
2352-3441