Determination of zinc (II) in pharmaceuticals based on a flow-through bulk optode
A method based on flow injection (FI), was applied for the determination of Zn (II) using a flow-through bulk optode membrane that incorporates 1-(2-pyridylazo)-2-naphthol in a plasticized poly (vinyl chloride) membrane entrapped in a cellulose support. The calibration graph plotting the reflectance...
Saved in:
Published in | Journal of pharmaceutical and biomedical analysis Vol. 29; no. 5; pp. 779 - 786 |
---|---|
Main Authors | , , , , |
Format | Journal Article |
Language | English |
Published |
Amsterdam
Elsevier B.V
31.07.2002
Elsevier Science |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | A method based on flow injection (FI), was applied for the determination of Zn (II) using a flow-through bulk optode membrane that incorporates 1-(2-pyridylazo)-2-naphthol in a plasticized poly (vinyl chloride) membrane entrapped in a cellulose support. The calibration graph plotting the reflectance at 562 nm versus [Zn (II)] was linear in the range 0.16–3.27 mg
l
−1 (2.5×10
−6–5×10
−5 M) with a detection limit of 0.10 mg
l
−1. The variation coefficients of the sensor response for 0.33 mg
l
−1 of Zn (II) were ±0.11% for consecutive measurements (
n=10), ±0.19% between days (
n=5) and ±0.22% between different membranes (
n=6). The sensor can be readily regenerated with the same acetic/acetate carrier of pH 3.9. The FI method proposed was applied to the determination of zinc (II) in pharmaceuticals. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 0731-7085 1873-264X |
DOI: | 10.1016/S0731-7085(02)00173-5 |