Antimicrobial efficacy of endoscopic disinfection procedures: a controlled, multifactorial investigation

Background: Adequate disinfection of endoscopes is essential to prevent environmental and patient-to-patient transmission of infectious agents, but data from controlled studies are limited. Moreover, there is controversy regarding current guidelines for disinfection. We compared the antimicrobial ef...

Full description

Saved in:
Bibliographic Details
Published inGastrointestinal endoscopy Vol. 50; no. 2; pp. 152 - 158
Main Authors Cronmiller, James R., Nelson, Daniel K., Salman, Ghassan, Jackson, Dana K., Dean, Robert S., Hsu, Joseph J., Kim, Chung H.
Format Journal Article
LanguageEnglish
Published New York, NY Mosby, Inc 01.08.1999
Elsevier
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Background: Adequate disinfection of endoscopes is essential to prevent environmental and patient-to-patient transmission of infectious agents, but data from controlled studies are limited. Moreover, there is controversy regarding current guidelines for disinfection. We compared the antimicrobial efficacy of several endoscopic disinfection procedures controlling for multiple factors that affect reprocessing. Methods: A colonoscope was contaminated with 108 CFU/mL of Enterococcus faecalis as a standardized inoculum. The colonoscope was passed through 1 of 16 study arms (5 reps/arm for a total of 80 runs) that were controlled for all possible combinations of the following variables: manual precleaning; 10-, 20-, or 45-minute glutaraldehyde exposure; air or ethanol drying; or automated reprocessing with peracetic acid (liquid sterilization system). Suction accessory channels and air-water channels were harvested for microbiologic culture. Results: Control runs (no cleaning or disinfection) recovered more than 5 × 107 CFU/mL from each sampling site. When each processing variable was isolated independent of other variables, the benefits of manual precleaning, longer soak times, and ethanol drying were apparent. When factors were combined, manual precleaning followed by 20- and 45-minute glutaraldehyde exposure and ethanol drying removed all test organisms, as did processing with the liquid sterilization system. Conclusion: Although the initial cost is higher, the automated liquid sterilization system provides effective sterilization and minimizes worker exposure. In units where chemical disinfection is used, our results suggest that manual precleaning followed by at least 20-minute glutaraldehyde exposure and ethanol rinse drying are sufficient to achieve complete disinfection. (Gastrointest Endosc 1999;50:152-8.)
Bibliography:ObjectType-Article-2
SourceType-Scholarly Journals-1
ObjectType-Feature-1
content type line 23
ISSN:0016-5107
1097-6779
DOI:10.1016/S0016-5107(99)70217-8