Patient‐maintained propofol sedation using reaction time monitoring: a volunteer safety study

Summary Previous volunteer studies of an effect‐site controlled, patient‐maintained sedation system using propofol have demonstrated a risk of over‐sedation. We have incorporated a reaction‐time monitor into the handset of the patient‐maintained sedation system to add an individualised patient‐feedb...

Full description

Saved in:
Bibliographic Details
Published inAnaesthesia Vol. 68; no. 2; pp. 154 - 158
Main Authors Allam, S., Anderson, K. J., O’Brien, C., Macpherson, J. A., Gambhir, S., Leitch, J. A., Kenny, G. N. C.
Format Journal Article
LanguageEnglish
Published Oxford, UK Blackwell Publishing Ltd 01.02.2013
Wiley-Blackwell
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Summary Previous volunteer studies of an effect‐site controlled, patient‐maintained sedation system using propofol have demonstrated a risk of over‐sedation. We have incorporated a reaction‐time monitor into the handset of the patient‐maintained sedation system to add an individualised patient‐feedback mechanism. This study assessed if such reaction‐time feedback modification would reduce the risk of over‐sedation in 20 healthy volunteers deliberately attempting to over‐administer themselves propofol. All the volunteers successfully sedated themselves without reaching any unsafe endpoints. All volunteers maintained verbal contact throughout, in accordance with the definition of conscious sedation. The mean (SD) lowest SpO2 was 97 (1.7) % when breathing room air and no volunteer required supplementary oxygen. The mean (SD) maximum effect‐site propofol concentration reached was 1.7 (0.4) μg.ml−1. The present system was found to be safer than its predecessors, allowing conscious sedation, but preventing over‐sedation.
Bibliography:Presented in part to the UK Society for Intravenous Anaesthesia, Chepstow, UK, 2006.
http://www.anaesthesiacorrespondence.com
You can respond to this article at
ObjectType-Article-2
SourceType-Scholarly Journals-1
ObjectType-Feature-1
content type line 23
ObjectType-Article-1
ObjectType-Feature-2
ISSN:0003-2409
1365-2044
DOI:10.1111/anae.12036