Dose to radiation therapists from activation at high-energy accelerators used for conventional and intensity-modulated radiation therapy

The increased beam-on times which characterize intensity-modulated radiation therapy (IMRT) could lead to an increase in the dose received by radiation therapists due to induced activity. To examine this, gamma ray spectrometry was used to identify the major isotopes responsible for activation at a...

Full description

Saved in:
Bibliographic Details
Published inMedical physics (Lancaster) Vol. 29; no. 4; p. 598
Main Authors Rawlinson, J Alan, Islam, Mohammad K, Galbraith, Duncan M
Format Journal Article
LanguageEnglish
Published United States 01.04.2002
Subjects
Online AccessGet more information

Cover

Loading…
More Information
Summary:The increased beam-on times which characterize intensity-modulated radiation therapy (IMRT) could lead to an increase in the dose received by radiation therapists due to induced activity. To examine this, gamma ray spectrometry was used to identify the major isotopes responsible for activation at a representative location in the treatment room of an 18 MV accelerator (Varian Clinac 21EX). These were found to be 28Al, 56Mn, and 24Na. The decay of the dose rate measured at this location following irradiation was analyzed in terms of the known half-lives to yield saturation dose rates of 9.6, 12.4, and 6.2 microSv/h, respectively. A formalism was developed to estimate activation dose (microSv/week) due to successive patient irradiation cycles, characterized by the number of 18 MV fractions per week, F, the number of MU per fraction, M, the in-room time between fractions, td (min), and the treatment delivery time t'r (min). The results are represented by the sum of two formulas, one for the dose from 28Al 1.8 x 10(-3) F M (1-e(-03t'(r))/t'r and one for the dose from the other isotopes approximately 1.1 x 10(-6) F(1.7) Mt(d). For conventional therapy doses are about 60 microSv/week for an 18 MV workload of 60,000 MU/week. Irradiation for QA purposes can significantly increase the dose. For IMRT as currently practiced, lengthy treatment delivery times limit the number of fractions that can be delivered per week and hence limit the dose to values similar to those in conventional therapy. However for an IMRT regime designed to maximize patient throughput, doses up to 330 microSv/week could be expected. To reduce dose it is recommended that IMRT treatments should be delivered at energies lower than 18 MV, that in multienergy IMRT, high-energy treatments should be scheduled in the latter part of the day, and that equipment manufacturers should strive to minimize activation in the design of high-energy accelerators.
ISSN:0094-2405
DOI:10.1118/1.1463063