We report on the first-ever use of a quantum ac source to calibrate a thermal transfer standard as part of the NIST calibration service for such devices, with reductions in calibration uncertainty of as much as an order of magnitude over traditional ac-dc transfer methods. We briefly describe the ba...

Full description

Saved in:
Bibliographic Details
Published inMetrologia Vol. 45; no. 3; pp. 275 - 280
Main Authors Lipe, Thomas E, Kinard, Joseph R, Tang, Yi-Hua, Benz, Samuel P, Burroughs, Charles J, Dresselhaus, Paul D
Format Journal Article
LanguageEnglish
Published Bristol IOP Publishing 01.06.2008
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:We report on the first-ever use of a quantum ac source to calibrate a thermal transfer standard as part of the NIST calibration service for such devices, with reductions in calibration uncertainty of as much as an order of magnitude over traditional ac-dc transfer methods. We briefly describe the basic quantum ac calibration system, its operation and measurement results. An analysis of the uncertainties for the measurements is presented and plans for further development are described. [PUBLICATION ABSTRACT]