Intra-abdominal pressure during endoscopic full-thickness resection comparing manual and automatic control insufflation: a block-randomized porcine study
Background and study aims An automatic carbon dioxide (CO 2 ) insufflating system (SPACE) was developed to stabilize intra-lumenal pressure (ILP) during endoscopic interventions. This study investigated whether SPACE could improve the control and monitoring of extra-lumenal intra-abdominal pressure...
Saved in:
Published in | Surgical endoscopy Vol. 34; no. 4; pp. 1625 - 1633 |
---|---|
Main Authors | , , , , , , , , , , , , |
Format | Journal Article |
Language | English |
Published |
New York
Springer US
01.04.2020
Springer Nature B.V |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Background and study aims
An automatic carbon dioxide (CO
2
) insufflating system (SPACE) was developed to stabilize intra-lumenal pressure (ILP) during endoscopic interventions. This study investigated whether SPACE could improve the control and monitoring of extra-lumenal intra-abdominal pressure (IAP) after establishing a perforation during endoscopic full-thickness resection (EFTR) of the gastric wall in porcine models.
Materials and methods
After first establishing the optimal preset pressure for gastric EFTR in four pigs, we compared IAP dynamics during EFTR between manual insufflation and SPACE using a block-randomized study (
n
= 10). IAP was percutaneously monitored and plotted on a timeline graph every 5 s. The maximal IAP and the area under the IAP curve exceeding 10 mmHg (AUC
≥
10 mmHg
) were compared between groups, with the agreement between IAP and endolumenally monitored ILP also analyzed for animals in the SPACE group.
Results
In the first study, 8 mmHg was identified as the most preferable preset pressure after establishment of the perforation. In the randomized study, the mean maximal IAP in the SPACE group was significantly lower than that in the manual insufflation group (11.0 ± 2.0 mmHg vs. 17.0 ± 3.5 mmHg;
P
= 0.03). The mean AUC
≥
10 mmHg
was also significantly smaller in the SPACE group. Bland–Altman analysis demonstrated agreement between IAP and ILP within a range of ± 1.0 mmHg.
Conclusions
SPACE could be used to control and safely monitor IAP during gastric EFTR by measuring ILP during perforation of the gastric wall. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 0930-2794 1432-2218 |
DOI: | 10.1007/s00464-019-06927-3 |