Establishment of Reference Intervals and Transfusion Criterion for Sonoclot Analysis

Sonoclot analyzer has been widely used in many countries. But the reference intervals provided by the manufacturer were derived from only 45 participants, and there was no cut-off value for transfusion for Sonoclot analysis. This study aimed to establish reference intervals and transfusion criterion...

Full description

Saved in:
Bibliographic Details
Published inJournal of Huazhong University of Science and Technology. Medical sciences Vol. 36; no. 4; pp. 614 - 617
Main Author 张真路 陈佑平 陶翠华 刘晓辉 李梦雅 周新
Format Journal Article
LanguageEnglish
Published Wuhan Huazhong University of Science and Technology 01.08.2016
Department of Laboratory Medicine, Zhongnan Hospital of Wuhan University, Wuhan 430071, China%Department of Pathology,Wuhan Asia Heart Hospital of Wuhan University, Wuhan 430022, China%Department of Transfusion,Wuhan Asia Heart Hospital of Wuhan University, Wuhan 430022, China%Department of Laboratory Medicine,Wuhan Asia Heart Hospital of Wuhan University, Wuhan 430022, China%Department of Cardiology, Wuhan Asia Heart Hospital of Wuhan University, Wuhan 430022, China
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Sonoclot analyzer has been widely used in many countries. But the reference intervals provided by the manufacturer were derived from only 45 participants, and there was no cut-off value for transfusion for Sonoclot analysis. This study aimed to establish reference intervals and transfusion criterion for Sonoclot analysis. Volunteers were recruited from healthy Chinese adults and patients undergoing cardiac surgery. Blood samples were withdrawn from forearm vein and measured for activated clotting time(ACT), clot rate(CR), platelet function(PF), activated partial thromboplastin time(APTT), fibrinogen concentration(FIB), and platelet count(PLT). The reference intervals were determined by the nonparametric method. Cut-off values were determined by the receiver operating characteristics curve. A total of 135 healthy volunteers and 281 patients were enrolled. The 95% reference intervals were 96–195 s, 22–51 signal U/min, 〉1.6 for ACT, CR, PF respectively. In the 281 patients, the results of APTT, FIB, PLT, ACT, CR, and PF ranged from 20.5–300.0 s, 0.28–4.11 g/L,(19.0–387.3)×109/L, 80–514 s, 2.9–74 signal U/min, and 0.1–5.1 respectively. The cut-off values for transfusion were 〉208, ≤14, and ≤1.3 for ACT, CR, PF respectively. The cut-off values of Sonoclot analysis were within the manufacturer's reference intervals, while they were outside the reference intervals established in this study. The results suggested that the manufacturer's reference intervals were not suitable for Chinese. The reference intervals and cut-off values established in this study will be helpful to Chinese patients.
Bibliography:Sonoclot analyzer reference interval transfusion criterion
42-1679/R
Sonoclot analyzer has been widely used in many countries. But the reference intervals provided by the manufacturer were derived from only 45 participants, and there was no cut-off value for transfusion for Sonoclot analysis. This study aimed to establish reference intervals and transfusion criterion for Sonoclot analysis. Volunteers were recruited from healthy Chinese adults and patients undergoing cardiac surgery. Blood samples were withdrawn from forearm vein and measured for activated clotting time(ACT), clot rate(CR), platelet function(PF), activated partial thromboplastin time(APTT), fibrinogen concentration(FIB), and platelet count(PLT). The reference intervals were determined by the nonparametric method. Cut-off values were determined by the receiver operating characteristics curve. A total of 135 healthy volunteers and 281 patients were enrolled. The 95% reference intervals were 96–195 s, 22–51 signal U/min, 〉1.6 for ACT, CR, PF respectively. In the 281 patients, the results of APTT, FIB, PLT, ACT, CR, and PF ranged from 20.5–300.0 s, 0.28–4.11 g/L,(19.0–387.3)×109/L, 80–514 s, 2.9–74 signal U/min, and 0.1–5.1 respectively. The cut-off values for transfusion were 〉208, ≤14, and ≤1.3 for ACT, CR, PF respectively. The cut-off values of Sonoclot analysis were within the manufacturer's reference intervals, while they were outside the reference intervals established in this study. The results suggested that the manufacturer's reference intervals were not suitable for Chinese. The reference intervals and cut-off values established in this study will be helpful to Chinese patients.
Zhen-lu ZHANG, You-ping CHEN, Cui-hua TAO , Xiao-hui LIU, Meng-ya LI, Xin ZHOU (1Department of Laboratory Medicine, Zhongnan Hospital of Wuhan University, Wuhan 430071, China; 2Department of Pathology, 3Department of Transfusion, 4Department of Laboratory Medicine, SDepartment of Cardiology, Wuhan Asia Heart Hospital of Wuhan University, Wuhan 430022, China)
ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:1672-0733
1993-1352
1993-1352
DOI:10.1007/s11596-016-1634-3