Automated oxygen control with O2matic ® during admission with exacerbation of COPD
It is a challenge to control oxygen saturation (SpO ) in patients with exacerbations of COPD during admission. We tested a newly developed closed-loop system, O2matic , and its ability to keep SpO within a specified interval compared with manual control by nursing staff. We conducted a crossover tri...
Saved in:
Published in | International journal of chronic obstructive pulmonary disease Vol. 13; pp. 3997 - 4003 |
---|---|
Main Authors | , , , , , |
Format | Journal Article |
Language | English |
Published |
New Zealand
Dove Medical Press Ltd
01.01.2018
Dove Medical Press |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | It is a challenge to control oxygen saturation (SpO
) in patients with exacerbations of COPD during admission. We tested a newly developed closed-loop system, O2matic
, and its ability to keep SpO
within a specified interval compared with manual control by nursing staff.
We conducted a crossover trial with patients admitted with an exacerbation of COPD and hypoxemia (SpO
≤88% on room air). Patients were monitored with continuous measurement of SpO
. In random order, they had 4 hours with manually controlled oxygen and 4 hours with oxygen delivery controlled by O2matic. Primary outcome was time within a prespecified SpO
target interval. Secondary outcomes were time with SpO
<85%, time with SpO
below target but not <85%, and time with SpO
above target.
Twenty patients were randomized and 19 completed the study. Mean age was 72.4 years and mean FEV
was 0.72 L (33% of predicted). Patients with O2matic-controlled treatment were within the SpO
target interval in 85.1% of the time vs 46.6% with manually controlled treatment (
<0.001). Time with SpO
<85% was 1.3% with O2matic and 17.9% with manual control (
=0.01). Time with SpO
below target but not <85% was 9.0% with O2matic and 25.0% with manual control (
=0.002). Time with SpO
above target was not significantly different between treatments (4.6% vs 10.5%,
=0.2). Patients expressed high confidence and a sense of safety with automatic oxygen delivery.
O2matic was able to effectively control SpO
for patients admitted with an exacerbation of COPD. O2matic was significantly better than manual control to maintain SpO
within target interval and to reduce time with unintended hypoxemia. |
---|---|
Bibliography: | ObjectType-Article-2 SourceType-Scholarly Journals-1 ObjectType-Feature-1 content type line 23 ObjectType-Undefined-3 |
ISSN: | 1178-2005 1176-9106 1178-2005 |
DOI: | 10.2147/COPD.S183762 |