A Membrane Ultrafiltration Procedure for Determining Diffusible Calcium in Serum
Abstract A rapid, simple procedure is described for separating diffusible from proteinbound serum calcium by centrifugation through high-flux ultrafiltration membranes. The mean normal value for ultrafilterable calcium was 57.1% (range 52 to 61%) of the total serum calcium concentration.
Saved in:
Published in | Clinical chemistry (Baltimore, Md.) Vol. 16; no. 3; pp. 226 - 228 |
---|---|
Main Authors | , , |
Format | Journal Article |
Language | English |
Published |
England
Am Assoc Clin Chem
01.03.1970
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Abstract
A rapid, simple procedure is described for separating diffusible from proteinbound serum calcium by centrifugation through high-flux ultrafiltration membranes. The mean normal value for ultrafilterable calcium was 57.1% (range 52 to 61%) of the total serum calcium concentration. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 0009-9147 1530-8561 |
DOI: | 10.1093/clinchem/16.3.226 |