HeartMate III: pump design for a centrifugal LVAD with a magnetically levitated rotor
A long-term, compact left ventricular assist device (LVAD), the HeartMate III, has been designed and fabricated, featuring a centrifugal pump with a magnetically levitated rotor. The pump has been optimized by in vitro testing to achieve a design point of 7 L/min against 135 mm Hg at high hydrodynam...
Saved in:
Published in | ASAIO journal (1992) Vol. 47; no. 4; p. 401 |
---|---|
Main Authors | , , , , , , , , , , , , , |
Format | Journal Article |
Language | English |
Published |
United States
01.07.2001
|
Subjects | |
Online Access | Get more information |
Cover
Loading…
Summary: | A long-term, compact left ventricular assist device (LVAD), the HeartMate III, has been designed and fabricated, featuring a centrifugal pump with a magnetically levitated rotor. The pump has been optimized by in vitro testing to achieve a design point of 7 L/min against 135 mm Hg at high hydrodynamic efficiency (30%) and to be capable of up to 10 L/min under such a load. Furthermore, the pump has demonstrated no mechanical failures, low hemolysis (4-10 mg/dl plasma free Hb), and low thrombogenicity during six (40, 27, 59, 42, 27, and 49-day) in vivo bovine studies. |
---|---|
ISSN: | 1058-2916 |
DOI: | 10.1097/00002480-200107000-00020 |