Evaluation of the clinical efficacy of a TW3-based fully automated bone age assessment system using deep neural networks

The aim of this study was to evaluate the clinical efficacy of a Tanner-Whitehouse 3 (TW3)-based fully automated bone age assessment system on hand-wrist radiographs of Korean children and adolescents. Hand-wrist radiographs of 80 subjects (40 boys and 40 girls, 7-15 years of age) were collected. Th...

Full description

Saved in:
Bibliographic Details
Published inImaging science in dentistry Vol. 50; no. 3; pp. 237 - 243
Main Authors Shin, Nan-Young, Lee, Byoung-Dai, Kang, Ju-Hee, Kim, Hye-Rin, Oh, Dong Hyo, Lee, Byung Il, Kim, Sung Hyun, Lee, Mu Sook, Heo, Min-Suk
Format Journal Article
LanguageEnglish
Published Korea (South) Korean Academy of Oral and Maxillofacial Radiology 01.09.2020
대한영상치의학회
Subjects
Online AccessGet full text
ISSN2233-7822
2233-7830
DOI10.5624/isd.2020.50.3.237

Cover

More Information
Summary:The aim of this study was to evaluate the clinical efficacy of a Tanner-Whitehouse 3 (TW3)-based fully automated bone age assessment system on hand-wrist radiographs of Korean children and adolescents. Hand-wrist radiographs of 80 subjects (40 boys and 40 girls, 7-15 years of age) were collected. The clinical efficacy was evaluated by comparing the bone ages that were determined using the system with those from the reference standard produced by 2 oral and maxillofacial radiologists. Comparisons were conducted using the paired -test and simple regression analysis. The bone ages estimated with this bone age assessment system were not significantly different from those obtained with the reference standard ( >0.05) and satisfied the equivalence criterion of 0.6 years within the 95% confidence interval (- 0.07 to 0.22), demonstrating excellent performance of the system. Similarly, in the comparisons of gender subgroups, no significant difference in bone age between the values produced by the system and the reference standard was observed ( >0.05 for both boys and girls). The determination coefficients obtained via regression analysis were 0.962, 0.945, and 0.952 for boys, girls, and overall, respectively ( =0.000); hence, the radiologist-determined bone ages and the system-determined bone ages were strongly correlated. This TW3-based system can be effectively used for bone age assessment based on hand-wrist radiographs of Korean children and adolescents.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
Nan Young Shin and Byoung-Dai Lee contributed equally to this work as the first authors.
ISSN:2233-7822
2233-7830