Fluorogenic Hand-Held Immunoassay for the Identification of Ricin: Rapid Analyte Measurement Platform
Rapid analyte measurement platform (RAMP™) fluorogenic hand-held immunoassays (HHAs) were evaluated for inclusivity/sensitivity, exclusivity/specificity, sample matrix effects, ruggedness/stability, and reproducibility in detection of ricin (RCA 60 ), a potential biological threat agent. The limit o...
Saved in:
Published in | Journal of immunoassay & immunochemistry Vol. 28; no. 3; pp. 227 - 241 |
---|---|
Main Authors | , |
Format | Journal Article |
Language | English |
Published |
England
Taylor & Francis Group
01.01.2007
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Rapid analyte measurement platform (RAMP™) fluorogenic hand-held immunoassays (HHAs) were evaluated for inclusivity/sensitivity, exclusivity/specificity, sample matrix effects, ruggedness/stability, and reproducibility in detection of ricin (RCA
60
), a potential biological threat agent. The limit of detection of HHAs for RCA
60
was 14 ng/mL or approximately 140 pg/test. HHAs were inclusive in detection of ricin RCA
60
, RCA
120
, ricin A chain, and ricin B chain and exclusive in discrimination of RCA
60
from other toxins. None of the sample matrices tested affected assay performance. RAMP™ ratios were tolerant of increases in sample volume, however, a decrease in sample volume of 25% resulted in significantly increased readings. RAMP™ readings were highly reproducible lot-to-lot, cartridge-to-cartridge, day-to-day, and reader-to-reader. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 1532-1819 1532-4230 |
DOI: | 10.1080/15321810701454730 |