Comparison of the Plusoptix A09 and Retinomax K-Plus 3 With Retinoscopy in Children
Purpose: To compare refractive measurements between the Plusoptix A09 (Plusoptix GmbH, Nuremberg, Germany), Retinomax K-Plus 3 (Right Med, Virginia Beach, VA), and retinoscopy in pediatric patients. Methods: Two hundred eyes of 200 patients (94 female, 106 male) were included in this cross-sectional...
Saved in:
Published in | Journal of pediatric ophthalmology and strabismus Vol. 52; no. 1; pp. 37 - 42 |
---|---|
Main Authors | , , , , , |
Format | Journal Article |
Language | English |
Published |
United States
SLACK INCORPORATED
01.01.2015
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Purpose:
To compare refractive measurements between the Plusoptix A09 (Plusoptix GmbH, Nuremberg, Germany), Retinomax K-Plus 3 (Right Med, Virginia Beach, VA), and retinoscopy in pediatric patients.
Methods:
Two hundred eyes of 200 patients (94 female, 106 male) were included in this cross-sectional study. Non-cycloplegic refraction was measured via the Plusoptix A09 and cycloplegic refraction was measured via the Retinomax K-Plus 3 and retinoscopy. Each device was evaluated for repeatability. The average measurements from all devices were compared.
Results:
The mean age of the patients was 6.2 ± 2.8 years (range: 4 to 12 years). The mean non-cycloplegic refraction via Plusoptix A09 was 0.74 ± 1.85 diopters (D) spherical and −0.79 ± 0.78 D cylindrical. The mean cycloplegic refraction via the Retinomax K-Plus 3 was 0.13 ± 1.92 D spherical and −0.74 ± 0.69 D cylindrical. The mean cyloplegic retinoscopy measurement was 0.71 ± 1.81 D spherical and −0.70 ± 0.76 D cylindrical. There was no significant difference between measurements (The mean age of the patients was 6.2 ± 2.8 years (range: 4 to 12 years). The mean non-cycloplegic refraction via Plusoptix A09 was 0.74 ± 1.85 diopters (D) spherical and −0.79 ± 0.78 D cylindrical. The mean cycloplegic refraction via the Retinomax K-Plus 3 was 0.13 ± 1.92 D spherical and −0.74 ± 0.69 D cylindrical. The mean cyloplegic retinoscopy measurement was 0.71 ± 1.81 D spherical and −0.70 ± 0.76 D cylindrical. There was no significant difference between measurements (
P
= .07).
Conclusions:
The Plusoptix A09 and Retinomax K-Plus 3 agree well with cycloplegic retinoscopy. Both devices may be used for screening children between 4 and 12 years old. The Plusoptix A09 may eliminate the need for cycloplegia for the detection of refractive errors in children.
[[
J Pediatr Ophthalmol Strabismus.
2015;52(1):37–42.] |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 0191-3913 1938-2405 |
DOI: | 10.3928/01913913-20141230-06 |