Comparison of the Plusoptix A09 and Retinomax K-Plus 3 With Retinoscopy in Children

Purpose: To compare refractive measurements between the Plusoptix A09 (Plusoptix GmbH, Nuremberg, Germany), Retinomax K-Plus 3 (Right Med, Virginia Beach, VA), and retinoscopy in pediatric patients. Methods: Two hundred eyes of 200 patients (94 female, 106 male) were included in this cross-sectional...

Full description

Saved in:
Bibliographic Details
Published inJournal of pediatric ophthalmology and strabismus Vol. 52; no. 1; pp. 37 - 42
Main Authors Yilmaz, Ihsan, Ozkaya, Abdullah, Alkin, Zeynep, Ozbengi, Sifa, Yazici, Ahmet Taylan, Demirok, Ahmet
Format Journal Article
LanguageEnglish
Published United States SLACK INCORPORATED 01.01.2015
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Purpose: To compare refractive measurements between the Plusoptix A09 (Plusoptix GmbH, Nuremberg, Germany), Retinomax K-Plus 3 (Right Med, Virginia Beach, VA), and retinoscopy in pediatric patients. Methods: Two hundred eyes of 200 patients (94 female, 106 male) were included in this cross-sectional study. Non-cycloplegic refraction was measured via the Plusoptix A09 and cycloplegic refraction was measured via the Retinomax K-Plus 3 and retinoscopy. Each device was evaluated for repeatability. The average measurements from all devices were compared. Results: The mean age of the patients was 6.2 ± 2.8 years (range: 4 to 12 years). The mean non-cycloplegic refraction via Plusoptix A09 was 0.74 ± 1.85 diopters (D) spherical and −0.79 ± 0.78 D cylindrical. The mean cycloplegic refraction via the Retinomax K-Plus 3 was 0.13 ± 1.92 D spherical and −0.74 ± 0.69 D cylindrical. The mean cyloplegic retinoscopy measurement was 0.71 ± 1.81 D spherical and −0.70 ± 0.76 D cylindrical. There was no significant difference between measurements (The mean age of the patients was 6.2 ± 2.8 years (range: 4 to 12 years). The mean non-cycloplegic refraction via Plusoptix A09 was 0.74 ± 1.85 diopters (D) spherical and −0.79 ± 0.78 D cylindrical. The mean cycloplegic refraction via the Retinomax K-Plus 3 was 0.13 ± 1.92 D spherical and −0.74 ± 0.69 D cylindrical. The mean cyloplegic retinoscopy measurement was 0.71 ± 1.81 D spherical and −0.70 ± 0.76 D cylindrical. There was no significant difference between measurements ( P = .07). Conclusions: The Plusoptix A09 and Retinomax K-Plus 3 agree well with cycloplegic retinoscopy. Both devices may be used for screening children between 4 and 12 years old. The Plusoptix A09 may eliminate the need for cycloplegia for the detection of refractive errors in children. [[ J Pediatr Ophthalmol Strabismus. 2015;52(1):37–42.]
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0191-3913
1938-2405
DOI:10.3928/01913913-20141230-06