A whole body counter for an emergency and occupational monitoring of an internal contamination with low energy photon emitters

A whole-body counter in SÚRO (NRPI) Prague, Czech Republic has been upgraded recently with the goal to enhance its capability of a safe, smooth, accurate and reproducible positioning of detectors for whole- and partial-body counting. The counter is intended especially for counting of low energy gamm...

Full description

Saved in:
Bibliographic Details
Published inRadiation physics and chemistry (Oxford, England : 1993) Vol. 116; pp. 160 - 164
Main Authors Fantinova, K, Fojtik, P, Pfeiferova, V
Format Journal Article
LanguageEnglish
Published Elsevier Ltd 01.11.2015
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:A whole-body counter in SÚRO (NRPI) Prague, Czech Republic has been upgraded recently with the goal to enhance its capability of a safe, smooth, accurate and reproducible positioning of detectors for whole- and partial-body counting. The counter is intended especially for counting of low energy gamma emitters in various organs and tissues of the human body. Counting efficiency calibration of a four-detector system installed in the shielded room has been performed by means of physical and voxel phantoms. The consistency of in vivo bioassay data of three internal contamination cases long-term monitored in the Institute is shown. •WBC chamber has been upgraded in order to enhance its counting capabilities.•Experimental and Monte Carlo technique were used for detector calibrations.•Consistency of the in vivo bioassay data for former and new WBC got to be preserved.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0969-806X
1879-0895
DOI:10.1016/j.radphyschem.2015.05.005