In Vivo Comparative Wear Study of Traditional and Highly Cross-linked Polyethylene in Total Hip Arthroplasty
In this study, we compare the in vivo wear performance of electron beam–irradiated, postirradiation-melted, highly cross-linked polyethylene (HXLPE) and traditional UHMWPE via the Martell method. Seventy hips with HXLPE performed at the Massachusetts General Hospital had 138 radiograph pairs for wea...
Saved in:
Published in | The Journal of arthroplasty Vol. 20; no. 7; pp. 880 - 886 |
---|---|
Main Authors | , , , , |
Format | Journal Article |
Language | English |
Published |
United States
Elsevier Inc
01.10.2005
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | In this study, we compare the in vivo wear performance of electron beam–irradiated, postirradiation-melted, highly cross-linked polyethylene (HXLPE) and traditional UHMWPE via the Martell method. Seventy hips with HXLPE performed at the Massachusetts General Hospital had 138 radiograph pairs for wear analysis and a 31.2-month average follow-up (range, 24-44 months). An age-matched, sex-matched, and body mass index–matched subgroup of 111 hips with 214 acceptable radiograph pairs and a 4-year follow-up from our previously published study on traditional polyethylene performed at Rush-Presbyterian–St. Luke's Medical Center served as a control group. Martell wear analysis was performed for each group. Overall and steady-state wear rates were compared via a specialized
t test. The steady-state wear in the HXLPE arm was observed after 2.0 years, was 0.007 mm/y, and was significantly less than the steady-state wear in the traditional arm (0.174 mm/y) (
P = .003). Highly cross-linked polyethylene penetration rate was not affected by sex, age, activity, or body mass index by Mann-Whitney analysis. |
---|---|