Real-World Patent Foramen Ovale (PFO) Closure in Japan ― 30-Day Clinical Outcomes From the AmplatzerTM PFO Occluder Japan Post-Marketing Surveillance Study

Background: The AmplatzerTM PFO Occluder was approved for marketing in Japan in May 2019, and the Amplatzer PFO Occluder Japan Post-marketing Surveillance (PFO Japan PMS) study was initiated in December 2019. This analysis presents 30-day clinical outcomes for PFO Japan PMS study patients.Methods an...

Full description

Saved in:
Bibliographic Details
Published inCirculation Journal Vol. 88; no. 9; pp. 1391 - 1397
Main Authors Akagi, Teiji, Hara, Hidehiko, Kanazawa, Hideaki, Fukui, Shigefumi, Hashimoto, Yoichiro, Iguchi, Yasuyuki, Iwama, Toru, Kataoka, Hiroharu, Kawamura, Akio, Kawano, Hiroyuki, Oki, Koichi, Yamagami, Hiroshi, the PFO Japan PMS Investigators
Format Journal Article
LanguageEnglish
Published The Japanese Circulation Society 23.08.2024
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Background: The AmplatzerTM PFO Occluder was approved for marketing in Japan in May 2019, and the Amplatzer PFO Occluder Japan Post-marketing Surveillance (PFO Japan PMS) study was initiated in December 2019. This analysis presents 30-day clinical outcomes for PFO Japan PMS study patients.Methods and Results: PFO Japan PMS is a prospective single-arm non-randomized multicenter clinical study. Eligible patients were indicated for patent foramen ovale (PFO) closure and underwent an implant attempt with the AmplatzerTM PFO Occluder. Technical success was defined as successful delivery and release of the occluder; procedural success was defined as technical success with no serious adverse events (SAEs) within 1 day of the procedure. The primary safety endpoint includes predefined device- and/or procedure-related SAEs through 30 days after the procedure. From December 2019 to July 2021, 500 patients were enrolled across 53 Japanese sites. The mean (±SD) patient age was 52.7±15.4 years, and 29.8% of patients were aged >60 years. Technical and procedural success rates were both high (99.8% and 98.8%, respectively). Further, there was only one primary safety endpoint event (0.2%): an episode of asymptomatic paroxysmal atrial fibrillation that occurred 26 days after the procedure.Conclusions: In this real-world Japanese study with almost one-third of patients aged >60 years, PFO closure with the AmplatzerTM PFO Occluder was performed successfully and safely, with a low incidence of procedure-related atrial arrhythmias.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:1346-9843
1347-4820
1347-4820
DOI:10.1253/circj.CJ-24-0080