Influence of decentration and tilt of Tecnis ZCB00 on visual acuity and higher order aberrations
Background/Objectives To determine the influence of decentration and tilt of a pseudophakic aspheric intraocular lens (IOL) on visual acuity (VA) and higher-order aberrations (HOAs), and to analyze the agreement between pupil center/axis and iridocorneal angles center/axis when assessing IOL decentr...
Saved in:
Published in | Eye (London) Vol. 37; no. 8; pp. 1640 - 1645 |
---|---|
Main Authors | , , , , , , , |
Format | Journal Article |
Language | English |
Published |
London
Nature Publishing Group UK
01.06.2023
Nature Publishing Group |
Subjects | |
Online Access | Get full text |
ISSN | 0950-222X 1476-5454 1476-5454 |
DOI | 10.1038/s41433-022-02211-2 |
Cover
Summary: | Background/Objectives
To determine the influence of decentration and tilt of a pseudophakic aspheric intraocular lens (IOL) on visual acuity (VA) and higher-order aberrations (HOAs), and to analyze the agreement between pupil center/axis and iridocorneal angles center/axis when assessing IOL decentration and tilt.
Subjects/Methods
A prospective interventional case series study including thirty-three patients undergoing Tecnis ZCB00 (Abbott Medical Optics) implantation. IOL decentration and tilt with respect to two reference systems (pupil and iridocorneal angles centers/axes), in cartesian (X,Y) and polar (radius/tilt, polar angle/azimuth) coordinates, were assessed with optical coherence tomography. VA and internal and ocular HOAs were evaluated. Multiple linear regression models and intraclass correlation coefficient (ICC) were computed.
Results
IOL decentration only showed a significant effect on internal HOAs for
Z
3
3
(
R
2
= 0.20,
P
= 0.04). IOL decentration with respect to the pupil center showed a significant effect on ocular
Z
3
−
3
(
R
2
= 0.18,
P
= 0.05),
Z
3
1
(
R
2
= 0.36,
P
= 0.001) and
Z
4
−
4
(
R
2
= 0.24,
P
= 0.02); and with respect to the center of iridocorneal angles, on ocular
Z
3
3
(
R
2
= 0.21,
P
= 0.03),
Z
4
2
(
R
2
= 0.32,
P
= 0.003), primary coma (
R
2
= 0.41,
P
< 0.001), and coma-like (
R
2
= 0.40,
P
= 0.001). Poor agreement between both reference systems was found for IOL decentration measurements (ICC ≤ 0.41), except for the polar angle coordinate (ICC = 0.83). Tilt measurements showed good agreement (ICC ≥ 0.75).
Conclusions
Tecnis ZCB00 decentration and tilt values after uneventful implantation appear not to have influence on VA, and their effect on HOAs are not high enough to clinically affect quality of vision. Pupil and iridocorneal angles used as reference systems may be interchangeable for IOL tilt measurements, but not for decentration. |
---|---|