Quantitative Testing of fMRI-Compatibility of an Electrically Active Mechatronic Device for Robot-Assisted Sensorimotor Protocols

Objective: To develop a quantitative set of methods for testing the functional magnetic resonance imaging (fMRI) compatibility of an electrically-active mechatronic device developed to support sensorimotor protocols during fMRI. Methods: The set of methods includes phantom and in vivo experiments to...

Full description

Saved in:
Bibliographic Details
Published inIEEE transactions on biomedical engineering Vol. 65; no. 7; pp. 1595 - 1606
Main Authors Farrens, Andria J., Zonnino, Andrea, Erwin, Andrew, O'Malley, Marcia K., Johnson, Curtis L., Ress, David, Sergi, Fabrizio
Format Journal Article
LanguageEnglish
Published United States IEEE 01.07.2018
The Institute of Electrical and Electronics Engineers, Inc. (IEEE)
Subjects
Online AccessGet full text
ISSN0018-9294
1558-2531
1558-2531
DOI10.1109/TBME.2017.2741346

Cover

More Information
Summary:Objective: To develop a quantitative set of methods for testing the functional magnetic resonance imaging (fMRI) compatibility of an electrically-active mechatronic device developed to support sensorimotor protocols during fMRI. Methods: The set of methods includes phantom and in vivo experiments to measure the effect of a progressively broader set of noise sources potentially introduced by the device. Phantom experiments measure the radio-frequency (RF) noise and temporal noise-to-signal ratio (tNSR) introduced by the device. The in vivo experiment assesses the effect of the device on measured brain activation for a human subject performing a representative sensorimotor task. The proposed protocol was validated via experiments using a 3T MRI scanner operated under nominal conditions and with the inclusion of an electrically-active mechatronic device - the MR-SoftWrist - as the equipment under test (EUT). Results: Quantitative analysis of RF noise data allows detection of active RF noise sources both in controlled RF noise conditions, and in conditions resembling improper filtering of the EUT's electrical signals. In conditions where no RF noise was detectable, the presence and operation of the EUT did not introduce any significant increase in tNSR. A quantitative analysis conducted on in vivo measurements of the number of active voxels in visual and motor areas further showed no significant difference between EUT and baseline conditions. Conclusion and significance: The proposed set of quantitative methods supports the development and troubleshooting of electrically-active mechatronic devices for use in sensorimotor protocols with fMRI, and may be used for future testing of such devices.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 14
content type line 23
ISSN:0018-9294
1558-2531
1558-2531
DOI:10.1109/TBME.2017.2741346