Physical activity monitoring during hospital stay: a validation study

The primary aim of this study was to investigate the concurrent validity of the PAM AM400 accelerometer for measuring physical activity in usual care in hospitalized patients by comparing it with the ActiGraph wGT3X-BT accelerometer. This was a prospective single centre observational study performed...

Full description

Saved in:
Bibliographic Details
Published inDisability and rehabilitation Vol. 45; no. 3; pp. 449 - 454
Main Authors Valkenet, Karin, Bor, Petra, Reijneveld, Elja, Veenhof, Cindy, Dronkers, Jaap
Format Journal Article
LanguageEnglish
Published England Taylor & Francis 30.01.2023
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:The primary aim of this study was to investigate the concurrent validity of the PAM AM400 accelerometer for measuring physical activity in usual care in hospitalized patients by comparing it with the ActiGraph wGT3X-BT accelerometer. This was a prospective single centre observational study performed at the University Medical Centre Utrecht in The Netherlands. Patients admitted to different clinical wards were included. Intraclass Correlation Coefficients (ICCs) were computed using a two-way mixed model with random subjects. Additionally, Bland-Altman plots were made to visualize the level of agreement of the PAM with the ActiGraph. To test for proportional bias, a regression analysis was performed. In total 17 patients from different clinical wards were included in the analyses. The level of agreement between the PAM and ActiGraph was found strong with an ICC of 0.955. The Bland-Altman analyses showed a mean difference of 1.12 min between the two accelerometers and no proportional bias (p = 0.511). The PAM is a suitable movement sensor to validly measure the active minutes of hospitalized patients. Implementation of this device in daily care might be helpful to change the immobility culture in hospitals. IMPLICATIONS FOR REHABILITATION Physical inactivity is common during hospital admission and the main cause of loss of muscle mass and physical fitness. The PAM AM400 is a suitable movement sensor to validly measure the active minutes of hospitalized patients. Implementation of this device in daily care might be helpful to change the immobility culture in hospitals.
ISSN:0963-8288
1464-5165
DOI:10.1080/09638288.2022.2034995